Fresenius 2008T Hemodialysis Machines Recalled for False USB Device Alarm
Fresenius Medical Care is recalling 2008T and 2008T Bluestar Hemodialysis Machines with CDX due to a software and hardware sensitivity to electromagnetic interference that triggers false alarms.
- Product
- 2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124…
- Category
- Medical Device
- Distribution
- Distributed nationwide