The Recall Desk

State

Mississippi product recalls

73,495 recalls have nationwide distribution and so reach Mississippi. 5,378 additional recalls listed Mississippi specifically in their distribution scope.

What the numbers show

Total
5,378
Critical
933
Severe
1,346
Most recent
2026-10-14

Of the 5,378 recalls naming Mississippi we have scored against our rubric, 933 (17%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1,346 (25%) are Severe.

These figures cover only the 5,378 recalls whose distribution notice names Mississippi specifically. The 73,495 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

22526–22550 of 78873

  • HighFDA (Devices)·Z-0912-2022·2022-04-20

    Orthopedic surgical hammertoe instrument recalled for improper component sizing

    Nextremity Solutions is recalling the Nextra Hammertoe Correction System due to an oversized driver component that makes it difficult to properly connect and remove during surgery.

    Product
    Nextra Hammertoe Correction System 3.5 Middle 3.2 Proximal Kit SKU: NX-3532K Orthopedic surgical instrument
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0987-2022·2022-04-20

    IHOP Pancake Mix Recalled Nationwide for Foreign Material Contamination

    Continental Mills is recalling 240 cases of IHOP Pancake Mix (Item #743-0278) nationwide due to the presence of foreign materials. Affected lots are KX2062 and KX2063 with a best-by date of 9/1/2022.

    Product
    Item # 743-0278, IHOP Pancake. Product is shelf stable and packaged in a box. net wt. 5 lbs., UPC 0 41449 47540 0.
    Category
    Food
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0900-2022·2022-04-20

    FDA Updates Labeling for LIPOSORBER LDL Adsorption Column Due to Anaphylactoid Shock Risk

    LIPOSORBER LA-15 LDL Adsorption Column labeling has been updated to address a risk of severe anaphylactoid reactions and shock in patients receiving treatment while taking ACE inhibitors.

    Product
    LIPOSORBER System, Model: LA-15, consisting of: Apheresis machine, product# 101445 model: MA-03, and disposables (product#, model) [#101447, KP-05 Sulflux Plasma Separator], [#101473, HDE KP-05 Sulflux Plasma Separator], [#101785, LA-15 (AU) LDL Adsorption Column…
    Category
    Medical Device
    Distribution
    34 states
  • HighFDA (Devices)·Z-0899-2022·2022-04-20

    iTotal Identity Knee Replacement System—incorrect tibial base plate packaged

    Conformis is recalling specific units of the iTotal Identity Posterior Stabilised Knee Replacement System due to incorrect tibial base plate implants being packaged in surgical kits.

    Product
    iTotal Identity Posterior Stabilised (PS) Knee Replacement System Model Number: TPS-301-1113-020101
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0939-2022·2022-04-20

    Cardiac electrode fails electrical safety test; defibrillation may not work

    The CLARAVUE Pre-wired Electrode does not meet electrical safety standards for defibrillators. It may not deliver effective electrical charge during emergency defibrillation.

    Product
    CLARAVUE Pre-wired Electrode - IP SET 6V RT ADULT-US Model: 32028780 50603-US
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0914-2022·2022-04-20

    Orthopedic surgical driver oversized, affecting hammertoe correction

    Nextremity Solutions is recalling the Nextra Hammertoe Correction System Driver (SKU: NX-DR) due to oversizing at the proximal end, which impairs connection and removal from the implant, causing surgical delays. The defect affects 239 units distributed in the US, Japan, and the Netherlands.

    Product
    Nextra Hammertoe Correction System Driver SKU: NX-DR Orthopedic surgical instrument
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0904-2022·2022-04-20

    AccessPLUS hemostasis valves recalled for internal guidewire entrapment risk

    Merit Medical is recalling AccessPLUS Large Bore Hemostasis Valves due to an internal design gap that can trap guidewires, potentially delaying medical procedures.

    Product
    AccessPLUS Large Bore Hemostasis Valves, Catalog Code: 580758001, 580774001, IN8152/C, K04-01197, K05T-01863, K09-09102B, K09-10388A, K09-12917, K09-13501AP, K10-05457AP, K10-05635, K10-05868P, K10-05872P, K12-08303, K12-10346, K12-10375, K12-10724AP, K12-10962, K12-11413…
    Category
    Medical Device
    Distribution
    49 states
  • HighFDA (Devices)·Z-0923-2022·2022-04-20

    Abbott Vascular Cardiovascular Device Recalled for Leaks and Loose Connections

    Abbott Vascular is recalling the 20/30 Priority Pack w/COPILOT device due to an increase in complaints of leaks and loose connections. The recall affects 131 lots distributed nationwide and internationally.

    Product
    20/30 Priority Pack w/COPILOT - REF 1003327 WPL2122272-01 (2019-03-01)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0986-2022·2022-04-20

    Krusteaz Pro Buttermilk Pancake Mix Recalled Nationwide for Foreign Materials

    Krusteaz Pro Buttermilk Pancake Mix is being recalled due to potential foreign materials. Approximately 970 cases with specific lot codes were distributed nationwide.

    Product
    Item # 731-0120, Krusteaz Pro Buttermilk Pancake Mix. Product is shelf stable and packaged in a box. net wt. 5 lbs., UPC 0 41449 10180 4.
    Category
    Food
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0906-2022·2022-04-20

    Integra Universal Flexible Arm Recall: Internal Cable May Break

    Integra LifeSciences is recalling the Universal Flexible Arm (part number REF 1362275) because excess tightening force can break the internal cable at the tip, rendering the device unusable.

    Product
    Integra Universal Flexible Arm part number REF 1362275
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0992-2022·2022-04-20

    Tattooed Chef Plant-Based Sausage Breakfast Bowl Recalled for False Gluten-Free Claim

    Ittella International recalls Tattooed Chef Plant-Based Sausage Breakfast Bowls nationwide due to an incorrect gluten-free claim on the container lid. The product contains wheat flour, which is correctly listed in the ingredients and allergen statement.

    Product
    Tattooed Chef Plant-Based Sausage Breakfast Bowl Frozen 18-month shelf life Box Net wt. 40oz (1.13kg) UPC: 850015982569 Packed for: Tattooed Chef Paramount, CA 90723
    Category
    Food
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0769-2022·2022-04-20

    Lansoprazole Delayed-Release Capsules Recalled for Out-of-Specification Dissolution

    Preferred Pharmaceuticals is recalling certain lots of Lansoprazole Delayed-Release Capsules 15mg due to dissolution testing results that do not meet specifications. Approximately 791 bottles were distributed nationwide.

    Product
    Preferred Pharmaceuticals, Inc. The Physician's Solution., Lansoprazole Delayed-Release Capsules, 15mg, Generic for Prevacid, 90-Count Bottle, Manufacturer: Dr Reddy's Laboratories Limited, NDC 68788-6390-09
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0924-2022·2022-04-20

    Abbott INDEFLATOR Plus 30 Device Recalled for Leaks and Connection Faults

    Abbott Vascular is recalling INDEFLATOR Plus 30 balloon inflators due to increasing complaints of leaks and intermittent loose connections that could affect device function during medical procedures.

    Product
    INDEFLATOR Plus 30 - REF 1000183 WPL2122277-01 (2019-03-01)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0905-2022·2022-04-20

    Breast Milk Analyzer Lacks FDA Validation, Risks Sub-optimal Infant Nutrition

    Stanbio Laboratory's Creamatocrit Plus breast milk analyzer was marketed without required FDA 510(k) clearance. Its performance has not been validated, and erroneous results could lead to inadequate nutritional supplementation for infants.

    Product
    Separation Technology Creamatocrit Plus, Catalog Number: 100-146
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0942-2022·2022-04-20

    ARIA Radiation Therapy Management Software May Cause Treatment Location Error

    Varian ARIA radiation therapy management software versions 13.6 through 16.1 may generate incorrect treatment plan values that could direct radiation therapy to the wrong location. No illnesses have been reported.

    Product
    ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0, 16.1
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0938-2022·2022-04-20

    Pre-wired ECG Electrode Set Fails Defibrillation Voltage Safety Standard

    Graphic Controls' CLARAVUE Pre-wired Electrode set does not meet required electrical specifications for defibrillation use. The device fails to withstand 5,000 volts without breakdown, potentially resulting in ineffective defibrillation.

    Product
    CLARAVUE Pre-wired Electrode - IP SET 12 RT ADULT-US Model: 32028779 50602-US
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0925-2022·2022-04-20

    Abbott Vascular Plus 30 Priority Pack Recalled for Connection Leaks

    Abbott Vascular is recalling seven lots of the Plus 30 Priority Pack due to increased reports of leaks and intermittent loose connections in the vascular devices.

    Product
    Plus 30 Priority Pack w/.096 RHV - REF 1000183 WPL2122277-01 (20-03-01) Plus 30 Priority Pack w/.115 RHV - REF 1000185-115 WPL2122279-01 (2019-03-01)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0913-2022·2022-04-20

    Hammertoe Correction Surgical System Recalled for Driver Connection Defect

    Nextremity Solutions recalls the Nextra Hammertoe Correction System because the driver's proximal end is oversized, making it difficult to connect and remove from the implant, potentially delaying surgery.

    Product
    Nextra Hammertoe Correction System 3.5 Middle and 3.2 Proximal Saw Cut SKU: NX-3532K-SC Orthopedic surgical instrument
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0989-2022·2022-04-20

    Cinnamon Crumb Cake Recalled for Potential Foreign Materials

    Continental Mills voluntarily recalls Krusteaz Cinnamon Crumb Cake (Lot KX2063) due to potential foreign materials that could pose a choking or injury risk.

    Product
    Item # 722-0105, Krusteaz Cinnamon Crumb Cake. Product is shelf stable and packaged in a box. net wt. 21 oz., UPC 0 41449 11140 7.
    Category
    Food
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0937-2022·2022-04-20

    CLARAVUE Pre-wired Electrode fails to meet electrical safety specification

    Graphic Controls' CLARAVUE Pre-wired Electrode units fail to meet electrical safety specifications required for defibrillation. The device delivers an inefficient electrical charge and may result in ineffective defibrillation.

    Product
    CLARAVUE Pre-wired Electrode - IP SET M5 RT ADULT-US Model: 32028778 50601-US
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0929-2022·2022-04-20

    Siemens Atellica CH Hemoglobin A1c Reagent Kits Recalled for Quality Control Issues

    Siemens Healthcare is recalling Atellica CH Hemoglobin A1c reagent kits due to potential reagent carryover affecting quality control and patient test results. The issue may cause inaccurate diabetes monitoring measurements.

    Product
    Atellica CH Enzymatic Hemoglobin A1c (A1c_E/A1c_H), Siemens Material Number 11097536
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0911-2022·2022-04-20

    Nextra Hammertoe Correction System surgical instruments recalled for defective connector

    Nextremity Solutions recalls 459 units of the Nextra Hammertoe Correction System due to an oversized proximal driver end that prevents proper connection and removal during surgery.

    Product
    Nextra Hammertoe Correction System 4.5 Middle and 3.2 Proximal Saw Cut SKU: NX-4532K-SC Orthopedic surgical instrument
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0984-2022·2022-04-20

    Great Value Buttermilk Pancake Mix Recalled for Foreign Materials

    Continental Mills is recalling Great Value Buttermilk Pancake & Waffle Mix (item #714-5000) due to foreign materials. Approximately 1,125 cases were distributed nationwide.

    Product
    Item # 714-5000, Great Value Buttermilk Pancake & Waffle Mix. Product is shelf stable and packaged in a box. net wt. 32 oz., UPC 0 78742 37082 8.
    Category
    Food
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-0766-2022·2022-04-20

    Cyanocobalamin Injection Recalled for Subpotent Drug - Out of Specification Assay

    Aurobindo Pharma USA Inc. is recalling 433,809 vials of Cyanocobalamin injection nationwide because assay testing showed the drug is subpotent, containing less than the specified 1,000 mcg/mL of active ingredient. Affected lots with specific expiration dates from September 2022 to April 2023 are identified for recall.

    Product
    CYANOCOBALAMIN — CYANOCOBALAMIN (CYANOCOBALAMIN)
    Category
    Drug
    Distribution
    Distributed nationwide