  • HighFDA (Devices)·Z-1750-2026·2026-04-08

    Paradigm REAL-Time Revel pump delivery varies with pump height changes

    Medtronic is recalling Paradigm REAL-Time Revel insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1696-2026·2026-04-08

    da Vinci monopolar curved scissors recalled over frayed grip cables

    Intuitive Surgical is recalling da Vinci S and Si Monopolar Curved Scissors Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1678-2026·2026-04-08

    Philips Azurion 7 M20 X-ray may not start from foot switch

    Philips has identified instances where X-ray imaging on the Azurion 7 M20 may not initiate, or may run intermittently, when the wired foot switch is used. 5,167 units are affected.

    Product
    Azurion 7 M20; System Code: (1)722079, (2)722224, (3)722234, (4)722282 (OUS only);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1706-2026·2026-04-08

    Medline bronchoscopy packs contain recalled Olympus biopsy valves shedding fragments

    Medline is recalling three bronchoscopy convenience kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1743-2026·2026-04-08

    MiniMed 640G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 640G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1752-2026·2026-04-08

    Paradigm REAL-Time Veo pump delivery varies with pump height changes

    Medtronic is recalling Paradigm REAL-Time Veo insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1689-2026·2026-04-08

    Jiffy composite polisher cups may crumble and break apart easily

    Ultradent Products is recalling one lot of Jiffy Original Composite Polisher Cups because the cups may crumble and break apart easily, which could delay patient treatment.

    Product
    Jiffy Original Composite Polisher Cups (Coarse) 12pk, REF 7011.
    Category
    Medical Device
    Distribution
    31 states
  • HighFDA (Drugs)·D-0410-2026·2026-04-08

    Quality Choice moisturizing relief eye drops recalled over sterility assurance

    Quality Choice Moisturizing Relief Eye Drops, containing dextran 70, polyethylene glycol 400, povidone and tetrahydrozoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    QUALITY CHOICE MOISTURIZING RELIEF EYE DROPS — QUALITY CHOICE MOISTURIZING RELIEF EYE DROPS (DEXTRAN 70, POLYETHYLENE GLYCOL 400, POVIDONE, TETRAHYDROZOLINE HCL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1766-2026·2026-04-08

    Oxoid Microbact Listeria kit may identify the incorrect organism

    Oxoid, part of Thermo Fisher Scientific, is recalling one lot of the Microbact 12L Listeria identification kit because it may identify the incorrect organism.

    Product
    OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1749-2026·2026-04-08

    Paradigm REAL-Time insulin pump delivery varies with pump height

    Medtronic is recalling Paradigm REAL-Time insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1747-2026·2026-04-08

    MiniMed 620G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 620G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 620G Insulin Pump (MMT-1750)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1710-2026·2026-04-08

    Medline bronchoscopy kits contain Olympus biopsy valves that shed fragments

    Medline is recalling six bronchoscopy and thoracic kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRON
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1708-2026·2026-04-08

    Medline endoscopy kits contain recalled Olympus biopsy valves shedding fragments

    Medline is recalling BAL and ENDO kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1690-2026·2026-04-08

    Medline ENFit G-tube connectors may leak from dimensional specification failure

    Medline is recalling ENFit G-tube connectors because they were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices.

    Product
    ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1691-2026·2026-04-08

    Medline ENFit G-tube connector kits may leak from dimensional defect

    Medline is recalling kits containing ENFit G-tube connectors because the connectors were not made to required dimensional specifications and may not form a proper seal.

    Product
    Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1701-2026·2026-04-08

    da Vinci permanent cautery hook recalled over frayed pitch cables

    Intuitive Surgical is recalling da Vinci S and Si Permanent Cautery Hook Instruments following increased complaints regarding frayed or broken pitch cables on reusable surgical instruments.

    Product
    Brand Name: da Vinci S, Si Permanent Cautery Hook Instrument Product Name: da Vinci S, Si Permanent Cautery Hook Instrument Model/Catalog Number: 420183 Software Version: NA Product Description: EndoWrist Instruments or da Vinci S/Si Reusable instruments, including blunt and
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0413-2026·2026-04-08

    GoodSense original formula eye drops recalled over sterility assurance

    GoodSense Eye Drops Original Formula, containing tetrahydrozoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    GOODSENSE EYE DROPS ORIGINAL FORMULA — GOODSENSE EYE DROPS ORIGINAL FORMULA (TETRAHYDROZOLINE HCL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1654-2026·2026-04-08

    Paradise renal denervation catheter unit distributed despite scrapping designation

    One Paradise Ultrasound Renal Denervation catheter was distributed although it was a nonconforming product that should have been scrapped and not used in humans.

    Product
    Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1753-2026·2026-04-08

    Burlington Medical thyroid shields may lose radiation attenuation over time

    Burlington Medical is recalling thyroid shields over the potential for attenuation degradation over time, which decreases the product's usable lifespan. 41,962 units are affected.

    Product
    Burlington Medical, Thyroid Shield.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1697-2026·2026-04-08

    da Vinci double fenestrated grasper recalled over broken grip cables

    Intuitive Surgical is recalling da Vinci S and Si Double Fenestrated Grasper Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci S, Si Double Fenestrated Grasper Instrument Model/Catalog Number: 420189 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including bl
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1761-2026·2026-04-08

    Burlington Medical leg wraps may lose radiation attenuation over time

    Burlington Medical is recalling leg wraps because of the potential for attenuation degradation over time, decreasing the lifespan of the product.

    Product
    Burlington Medical, Leg Wraps.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1698-2026·2026-04-08

    da Vinci mega needle driver recalled over broken or frayed cables

    Intuitive Surgical is recalling da Vinci S and Si Mega Needle Driver Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Mega Needle Driver Instrument Product Name: da Vinci S, Si Mega Needle Driver Instrument Model/Catalog Number: 420194 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp en
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0419-2026·2026-04-08

    Prazosin hydrochloride 100-count bottles recalled over nitrosamine impurity

    A further quantity of Prazosin Hydrochloride tablets is being recalled after detection of the nitrosamine impurity N-nitroso-prazosin impurity C above acceptable limits.

    Product
    PRAZOSIN HYDROCHLORIDE — PRAZOSIN HYDROCHLORIDE (PRAZOSIN HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1748-2026·2026-04-08

    Paradigm insulin pump delivery varies with pump height changes

    Medtronic is recalling Paradigm insulin pumps models MMT-712 and MMT-715 because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm Insulin Pump (MMT-712, MMT-715)
    Category
    Medical Device
    Distribution
    Distributed nationwide