  • HighFDA (Devices)·Z-1643-2026·2026-04-01

    Olympus PKS cutting forceps 920000PK jaws may break from welds

    Olympus is recalling PKS Cutting Forceps with cord, model 920000PK, because a supplier did not adequately validate the welding process and defective welds can cause the jaw to break.

    Product
    Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1605-2026·2026-04-01

    I.T.S. foot locking plates heat more in MRI than stated

    I.T.S. GmbH is recalling the FLS Foot Locking Plates System and HOL Hallux Osteotomy Locking Plate because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18, 373
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1603-2026·2026-04-01

    I.T.S. femur and tibia plates heat more in MRI than stated

    I.T.S. GmbH is recalling the LRS Locking Reconstruction System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL (Distal Femur Locking) Distal Femur Plate; Article Numbers: 21301-5, 21301-7, 21301-9, 21301-13, 21302-5, 21302-7, 21302-9, 21302-13. 2. PTL (Proximal Lateral
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1641-2026·2026-04-01

    Olympus HALO PKS cutting forceps jaws may break from defective welds

    Olympus is recalling HALO PKS Cutting Forceps because a supplier did not adequately validate the welding process, and defective welds can cause the forceps jaw to break during clinical use.

    Product
    Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electro
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1635-2026·2026-04-01

    Medtronic MiniMed 780G software defects may cause insulin delivery inaccuracies

    Medtronic is recalling MiniMed 780G insulin pumps on software 6.60 and 6.61 because three software defects could result in hypoglycemia or hyperglycemia through inaccurate insulin delivery.

    Product
    MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0406-2026·2026-04-01

    Teva octreotide acetate injection recalled over sterility assurance deficiencies

    Teva is recalling twelve lots of Octreotide Acetate for Injectable Suspension after quality system deficiencies were identified during an FDA inspection at the contract manufacturer.

    Product
    OCTREOTIDE ACETATE — OCTREOTIDE ACETATE (OCTREOTIDE ACETATE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1634-2026·2026-04-01

    TECNIS Eyhance intraocular lens haptic may stick to the optic

    AMO Puerto Rico Manufacturing is recalling TECNIS Eyhance IOLs with the TECNIS Simplicity Delivery System because a limited quantity may have a haptic that sticks to the optic, preventing consistent unfolding.

    Product
    Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1610-2026·2026-04-01

    Olympus endoscope suction valve no longer compatible with named reprocessors

    Olympus is recalling the MAJ-1443 Endoscope Suction Valve because the MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

    Product
    Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1595-2026·2026-04-01

    I.T.S. calcaneus repair plate heats more in MRI than stated

    I.T.S. GmbH is recalling the FR.O.H. Calcaneus Repair System anterolateral calcaneus plate because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1609-2026·2026-04-01

    Beckman Coulter MicroScan gram-negative panels may contain contaminated wells

    Beckman Coulter is recalling one lot of MicroScan Neg MIC 56 panels due to the likely presence of contamination in one or more wells.

    Product
    MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1632-2026·2026-04-01

    Straumann custom abutments made from wrong bar material after mix-up

    Straumann is recalling Custom Abutment Ti for Straumann RC because a material mix-up meant titanium abutments were manufactured from TAN bar material instead of Titanium Grade 4.

    Product
    Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1598-2026·2026-04-01

    I.T.S. straight and forearm plates heat more in MRI than stated

    I.T.S. GmbH is recalling Straight Plates with Angular Stability, including forearm and compression plates, because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use reflected.

    Product
    I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Pla
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1593-2026·2026-04-01

    I.T.S. clavicle plates heat more in MRI than instructions state

    I.T.S. GmbH is recalling Clavicula Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm; Article
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1602-2026·2026-04-01

    I.T.S. distal humeral plates heat more in MRI than stated

    I.T.S. GmbH is recalling Distal Humeral Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N, 373
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1613-2026·2026-04-01

    Olympus OER-Pro reprocessor no longer compatible with two valve models

    Olympus is recalling the OER-Pro Endoscope Reprocessor because the MAJ-1443 and MAJ-1444 air/water valves are no longer reprocessing compatible with the OER-Pro and OER-Elite reprocessors.

    Product
    Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0599-2026·2026-04-01

    Ajinomoto Green Gyoza Vegetable Edamame Dumplings Recalled for Glass

    Ajinomoto Foods North America Inc. is recalling Ajinomoto Green Gyoza Vegetable & Edamame Dumplings due to the presence of glass fragments (1–3 cm long and 2–4 mm wide) in the product.

    Product
    Item 6430093 Ajinomoto Green Gyoza Vegetable & Edamame Dumplings 12ct. Retail bag UPC: None. Product packaging is flexible bags in a master case or flexible bags in cartons in a master case. There are 20/12ct. bags per case. Case UPC 10071757080246. All products are intended to
    Category
    Food
    Distribution
    43 states
  • HighFDA (Food)·H-0601-2026·2026-04-01

    Ajinomoto Vegetable Gyoza Dumplings Recalled for Glass Contamination

    Ajinomoto Foods North America Inc. is recalling Ajinomoto Vegetable Gyoza Dumplings due to the presence of glass fragments in the product. The recalled dumplings were distributed nationwide.

    Product
    Item 6430153 Ajinomoto Vegetable Gyoza Dumplings. Net wt. 24.7oz. per bag. Retail box UPC: 071757010109. Product packaging is flexible bags in a master case or flexible bags in cartons in a master case. There are 8/24.7oz bags or 12.34lb. per case. Case UPC 10071757010106. All
    Category
    Food
    Distribution
    43 states
  • HighFDA (Devices)·Z-1638-2026·2026-04-01

    Olympus Everest bipolar forceps model 3005 jaws may break

    Olympus is recalling Everest Bipolar Cutting Forceps model 3005 because a supplier did not adequately validate the welding process, and defective welds can cause the jaw to break during clinical use.

    Product
    Brand Name: Olympus Everest Bipolar Cutting Forceps, 5 mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3005 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved u
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0594-2026·2026-04-01

    Ajinomoto Japanese-Style Vegetable Fried Rice Recalled for Glass Contamination

    Ajinomoto Foods North America Inc. is recalling Item 5650893 Japanese-Style Vegetable Fried Rice because some packages may contain glass fragments measuring 1–3 cm long and 2–4 mm wide.

    Product
    Item 5650893 Ajinomoto Japanese-Style Vegetable Fried Rice made with hijiki seaweed, fried tofu, edamame, and other colorful vegetables in an umami dashi sauce Net wt. 17oz. Retail box UPC 071757508934. Product packaging is flexible bags in a master case or flexible bags in cart
    Category
    Food
    Distribution
    43 states
  • HighFDA (Food)·H-0615-2026·2026-04-01

    Savannah Bee Company Honey BBQ Sauce Mustard Undeclared Allergens

    Savannah Bee Company is recalling Honey BBQ Sauce Mustard bottles containing an undeclared different flavor product with soy and wheat allergens not listed on the label.

    Product
    SAVANNAH BEE COMPANY HONEY BBQ SAUCE MUSTARD NET 16 FL OZ (453g), PACKED IN BOTTLES. INGREDIENTS: MUSTARD (DISTILLED VINEGAR, WATER, MUSTARD SEED, SALT, SPICES (INCLUDING PAPRIKA AND TUMERIC), NATURAL FLAVOR, GARLIC), HONEY, WATER, DISTILLED VINEGAR, SUGAR, TOMATO CONCENTRATE, HA
    Category
    Food
    Distribution
    45 states
  • HighFDA (Food)·H-0598-2026·2026-04-01

    Kroger Chinese Inspiration Vegetable Fried Rice recalled for glass

    Ajinomoto Foods North America Inc. is recalling Kroger Chinese Inspiration Vegetable Fried Rice due to the presence of glass fragments varying from 1–3 cm long and 2–4 mm wide.

    Product
    Item 81153 Kroger Chinese Inspiration Vegetable Fried Rice - Blend of long grain rice assorted vegetables & egg. Net wt. 22oz. Retail box UPC 011110041623. Product packaging is flexible bags in a master case or flexible bags in cartons in a master case. There are 6/22oz. boxes o
    Category
    Food
    Distribution
    43 states
  • HighFDA (Food)·H-0589-2026·2026-04-01

    Chilli Spanish Peanuts First Street recalled for glass contamination

    Western Mixers Produce & Nuts, Inc. is recalling First Street Chilli Spanish Peanuts (11 oz) because glass fragments were found in sealed containers. Affected lots have expiration dates of 2/11/2027, 2/13/2027, and 2/23/2027.

    Product
    Chilli Spanish Peanuts, First Street brand, 11oz clear plastic container. UPC: None Item Code: 2608-T20
    Category
    Food
    Distribution
    Distributed nationwide
  • HighNHTSA·26V209000·2026-04-01

    Audi Q3 recalled over remote sunroof closing from out of sight

    Volkswagen Group of America is recalling 6,077 Audi Q3 vehicles because incorrect software may allow remote sunroof closing from a distance where the operator cannot clearly see the sunroof.

    Product
    AUDI — AUDI Q3
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1592-2026·2026-04-01

    I.T.S. volar radius plates heat more in MRI than stated

    I.T.S. GmbH is recalling Volar Radius Plates with Angular Stability because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.

    Product
    Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3.
    Category
    Medical Device
    Distribution
    Distributed nationwide