The Recall Desk

State

Indiana product recalls

73,827 recalls have nationwide distribution and so reach Indiana. 9,598 additional recalls listed Indiana specifically in their distribution scope.

What the numbers show

Total
9,598
Critical
1,908
Severe
2,536
Most recent
2026-11-04

Of the 9,598 recalls naming Indiana we have scored against our rubric, 1,908 (20%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2,536 (26%) are Severe.

These figures cover only the 9,598 recalls whose distribution notice names Indiana specifically. The 73,827 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

51251–51275 of 83425

  • SevereFDA (Devices)·Z-0653-2017·2016-11-30

    AtriCure Recalls COBRA Fusion Ablation Systems Due to Component Separation Risk

    AtriCure is recalling COBRA Fusion Ablation Systems because a part may separate and remain in the body, potentially causing permanent impairment or damage.

    Product
    COBRA Fusion 50 Ablation System, COBRA Fusion 150 Ablation System, and COBRA Fusion Magnetic Retriever System. Model Numbers: 700-001, 700-001S, 700-001MI, 700-002, and 700-003. Intended to ablate cardiac tissue during cardiac surgery using radiofrequency (RF) energy when…
    Category
    Medical Device
    Distribution
    37 states
  • SevereFDA (Food)·F-0527-2017·2016-11-30

    Free2b Sun Cups Chocolate Recalled for Potential Milk Contamination

    Seth Ellis Chocolatier is recalling Free2b Sun Cups Chocolate due to potential contamination with milk ingredients. The recall covers 189 master cases distributed nationwide.

    Product
    Free2b Sun Cups Chocolate. Made With SunButter Sunflower Spread. Packaged in 2 Cups. Net weight 1.5 oz (45g) and 1 Cup. Net weight 0.75 oz (21g)
    Category
    Food
    Distribution
    Distributed nationwide
  • SevereFDA (Drugs)·D-0146-2017·2016-11-30

    VistaPharm Recalls Lactulose Solution Due to Bacterial Contamination

    VistaPharm is recalling Lactulose Solution, USP, 10 g/15 mL, because bulk solution tested positive for Burkholderia cepacia bacteria.

    Product
    Lactulose Solution, USP, 10 g/15 mL, 15 mL unit dose cups (NDC 66689-039-01), packaged in 15 mL x 50 unit dose cups per case (NDC 66689-039-50), Rx only, Manufactured by VistaPharm, Inc., Largo, FL 33771.
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0635-2017·2016-11-30

    Merit Medical Procedure Kits Recalled for Compromised Sterile Barrier

    Merit Medical Systems is recalling 53 Procedure Kits due to a compromised sterile barrier. The kits were distributed nationwide to specific states and Japan.

    Product
    Procedure Kit. Catalog Numbers K12-04830B and K12-YP030A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0655-2017·2016-11-30

    Hologic Recalls SURESOUND22 Uterine Sounds Due to Sterile Barrier Breach

    Hologic is recalling SURESOUND22 uterine sound multipacks because a breach in the sterile barrier compromises product sterility.

    Product
    SURESOUND22US Multipack (US Distribution) containing 2 packs of six devices. SURESOUND22 Multipack (OUS Distribution) containing 2 packs of six devices. For uterine sound
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0530-2017·2016-11-30

    EN-R-G Foods Recalls Protein Chews Due to Nausea and Vomiting Complaints

    EN-R-G Foods is recalling Cherry Lime, Raspberry, and Juneberry Protein Chews due to consumer complaints of nausea and vomiting after consumption.

    Product
    Protein Chews Raspberry,1.8oz, film package, 12 packets per box.
    Category
    Food
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0632-2017·2016-11-30

    Merit Medical Recalls Manifold Kits Due to Compromised Sterile Barrier

    Merit Medical Systems is recalling 40 Manifold Kits (Catalog No. K09-11867AP) because the sterile barrier is compromised, posing a risk of infection.

    Product
    Manifold Kit. Catalog Number K09-11867AP
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0634-2017·2016-11-30

    Merit Medical Recalls Pressure Monitoring Kits Due to Sterile Barrier Compromise

    Merit Medical Systems is voluntarily recalling 130 MeriTrans Pressure Monitoring Kits because a compromised sterile barrier may affect product sterility.

    Product
    Pressure Monitoring Kit (MeriTrans"). Catalog Number K11-00232G
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0622-2017·2016-11-30

    Toshiba Aquilion Prime CT System Recalled for Cooling Oil Leak Risk

    Toshiba American Medical Systems is recalling Aquilion Prime CT systems because cooling oil in the x-ray tube heat exchanger may leak during operation.

    Product
    Aquilion Prime TSX-303A Aquilion PRIME is a multislice helical CT system with an 40-row detector. High-speed rotation allows rapid data acquisition and shortens scan times, while the fast reconstruction unit further improves throughput, reducing the time required for diagnosis.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·16V861000·2016-11-30

    Cummins Recalls 2015-2016 Generator Sets Due to Door Hinge Failure Risk

    Cummins is recalling 164 model year 2015-2016 generator sets because door hinges may fail, causing doors to detach and create road hazards.

    Product
    CUMMINS POWER GENERATION — CUMMINS POWER GENERATION GENERATOR SET
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0375-2017·2016-11-30

    Philips CT Scanner Recall Due to Tabletop Movement Failure

    Philips is recalling 1,363 CT scanners because the patient support tabletop may fail to move horizontally during scans due to a mechanical linkage failure.

    Product
    Philips Healthcare System X-ray Tomography Computed :
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·16V863000·2016-11-30

    LGS Recalls 2017 Pace, Cargo Express, and Look Cargo Trailers

    LGS Industries is recalling 2017 Pace, Cargo Express, and Look cargo trailers because a defective weld could cause the ramp door bracket to detach, creating a crushing hazard.

    Product
    PACE — PACE, CARGO EXPRESS, LOOK CARGO TRAILER
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0633-2017·2016-11-30

    Merit Medical Procedure Kits Recalled for Compromised Sterile Barrier

    Merit Medical Systems is voluntarily recalling specific Procedure Kits due to a compromised sterile barrier, which may affect the sterility of the contents.

    Product
    Waste Management Kit. Catalog Number K10-04381AP
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0652-2017·2016-11-30

    Siemens Dimension Vista Chemistry 1 Calibrator Recall for Low Calcium Recovery

    Siemens is recalling Dimension Vista Chemistry 1 Calibrators (lots 5GM081 and 5GM082) because they may cause discrepant low calcium recovery in quality control and patient samples.

    Product
    Dimension Vista Chemistry 1 Calibrator The Chemistry 1 Calibrator (KC110B) is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA/CRE2), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0651-2017·2016-11-30

    Bausch + Lomb Recalls PeroxiClear Peroxide Solution for Excess Residuals

    Bausch + Lomb is recalling PeroxiClear Peroxide Solution because the product was found to contain excess residual levels of hydrogen peroxide.

    Product
    PeroxiClear Peroxide Solution, (3oz. and 12 oz., 12 oz. Twin Pack, 2x12 oz. Twin Pack for Target stores, 12 oz. for Walmart Vision Center and 3 oz. Starter Kit US), Manufactured by Bausch + Lomb, Rochester, NY 14609. PeroxiClear 3% Hydrogen Peroxide Cleaning & Disinfecting…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0645-2017·2016-11-30

    DiaSorin Recalls LIAISON Estradiol II Gen Assay Due to Cross-Reactivity

    DiaSorin is recalling the LIAISON Estradiol II Gen assay because it may produce inaccurate results for patients treated with fulvestrant.

    Product
    LIAISON Estradiol II Gen, 310680, 310680-CN The DiaSorin LIAISON Estradiol II Gen assay is a chemiluminescent immunoassay (CLIA) intended for the quantitative determination of estradiol in human serum. Assay results should be used in conjunction with other clinical and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0376-2017·2016-11-30

    Philips CT Scanner Recall Due to Potential Tabletop Movement Failure

    Philips is recalling 1,363 CT scanners because the patient support tabletop may fail to move horizontally during scans due to a mechanical linkage failure.

    Product
    Philips Healthcare System X-ray Tomography Computed :
    Category
    Medical Device
    Distribution
    Distributed nationwide