Most US product recalls have nationwide distribution — the manufacturer ships into every state and the recall sweeps the same way. We include those in this listing because they affect Illinois consumers even though the source notice doesn’t name the state. State-specific recalls are usually for products with regional distribution (a regional dairy, a local store brand) or where the agency was able to narrow the affected shipments to specific destinations.

If you’re checking whether a recall affects your purchase, the recall detail page lists the exact production lot codes, UPCs, and best-by dates. Match those against your package — distribution scope is a starting filter, not the final answer.

11151–11175 of 20304

  • HighFDA (Devices)·Z-0371-2024·2023-11-29

    Compress Device Anchor Plugs recalled due to metal burrs

    Biomet is recalling Compress Device Segmental Anchor Plugs due to metal burrs in transverse holes that may prevent drill or pins from passing through, potentially extending surgical time.

    Product
    Compress Device Segmental Anchor Plug, 12 MM -For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178402
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0325-2024·2023-11-29

    IRON assay kits recalled for inaccurate diagnostic results

    Sentinel CH SpA is recalling IRON assay kits (Reference Numbers 6K95-41 and 6K95-30) that may produce falsely elevated results. Affected lots could cause inaccurate iron measurements and delays in patient test reporting.

    Product
    IRON assay, Reference Numbers 6K95-41 and 6K95-30
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0397-2024·2023-11-29

    EXALT Model D Single-Use Duodenoscope Recalled for Lens Fluid Ingress

    Boston Scientific is recalling EXALT Model D Single-Use Duodenoscopes due to fluid ingress in the lens, which causes poor image quality during procedures.

    Product
    EXALT Model D Single-Use Duodenoscope, 2 pack: Outer box UPN# M00542421, Inner box UPN # M00542420 & Outer box UPN# M0054242CE1, Inner box UPN # M0054242CE0.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0350-2024·2023-11-29

    BD Gravity Infusion Sets Labeled DEHP-Free But Contain DEHP

    BD gravity blood infusion sets labeled as free of DEHP (di(2-ethylhexyl) phthalate) actually contain the substance. The recall affects products distributed worldwide through July 31, 2026.

    Product
    BD gravity sets (Product Name, Catalog #) 32-IN GRAV SET W/0.2 MICRON FILTER BAG ACCESS PORT SS VALVE TEXIUMTM LOW ABSORPTION, Catalog # 4030B-07T and BD GRAVITY BLOOD SET 15 DROP 180 MICRON FILTER SmartSiteTM VALVE NONVENTED, Catalog # 42081E.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0365-2024·2023-11-29

    SafeSpot Manual Resuscitators Recalled for CO2 Rebreathing Risk

    SunMed Holdings is recalling SafeSpot manual resuscitators with integrated manometers due to a backwards leak that allows CO2 rebreathing. The defect affects multiple accessory configurations distributed nationwide.

    Product
    Ventlab, LLC SafeSpot manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. SafeSpot Infant Resuscitator REF #s: SS3200OBPC-T, SS3204MB-D12, SS3200MB-2, SS3200MB-D1, SS3200OBP, SS3200MB-MMC, SS3200MBP-M00, SS3200MB-PW, SS3200MBP-2I, S
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0326-2024·2023-11-29

    Alinity c Iron Reagent Recalled for Falsely Elevated Test Results

    The FDA is recalling Alinity c Iron Reagent lots due to inaccurate results. The reagent produces falsely elevated patient iron values and may prevent proper quality control verification.

    Product
    Alinity c Iron Reagent, Reference Number 08P3920
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0373-2024·2023-11-29

    Biomet Compress Device Segmental Anchor Plugs Recalled for Metal Burrs

    Biomet has recalled certain Compress Device Segmental Anchor Plugs (16 MM) due to metal burrs in the transverse holes. These burrs may prevent the drill or pins from passing through, potentially extending surgical time.

    Product
    Compress Device Segmental Anchor Plug, 16 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178406
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0363-2024·2023-11-29

    Horizon manual resuscitators recalled for CO2 rebreathing hazard

    SunMed Holdings is recalling Ventlab Horizon manual resuscitators due to a backwards leak in the integrated manometer that allows carbon dioxide rebreathing. The affected devices were manufactured from September 2018 to the present and distributed nationwide.

    Product
    Ventlab, LLC Horizon manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Horizon Adult Resuscitator REF #s: 9122HZ, HZI9000, HZI9108, 9106HZ, 9110HZ, 9120HZ, 9111HZ, 9100HZ, 9105HZ, 9108HZ, 9006HZ, 9002HZ; 2. Horizon Infant Resu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0390-2024·2023-11-29

    Philips Achieva 1.5T MRI Device Recalled for Gradient Coil Fire Risk

    Philips is recalling 68 Achieva 1.5T MRI machines due to potential gradient coil component failure that could produce smoke and fire. No injuries have been reported.

    Product
    Achieva 1.5T Conversion
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0333-2024·2023-11-29

    Gastroscope Devices Recalled for Inadequate Channel Drying Process

    Aizu Olympus recalls 680 gastroscope-EUS devices due to unvalidated channel drying. Water remaining in the channels poses contamination and infection risks.

    Product
    Gastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for endoscopic surgery within the upper digestive tract. Model Numbers: GF-UC140P-AL5, GF-UCT180, GF-UE160-AL5.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0391-2024·2023-11-29

    Philips Achieva 1.5T MRI Systems Recalled for Gradient Coil Fire Risk

    Philips is recalling Achieva 1.5T MRI systems because the gradient coil component could fail and produce smoke or fire. The recall affects 2 units distributed worldwide.

    Product
    Achieva 1.5T Initial System
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0386-2024·2023-11-29

    Surgical implant anchor plugs may contain metal burrs affecting insertion

    Biomet's Compress Device Short Anchor Plugs may contain metal burrs in the transverse holes, which can prevent proper insertion of surgical tools and extend surgery time. The affected devices are distributed nationwide and internationally.

    Product
    Compress Device Short Anchor Plug, 24 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178566
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0348-2024·2023-11-29

    Uretero-reno fiberscopes recalled due to inadequate channel drying

    Aizu Olympus is recalling uretero-reno fiberscopes because the channel air drying process was not validated. A small percentage of repaired devices had residual water in the channels, creating a risk of contamination and infection.

    Product
    Uretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0331-2024·2023-11-29

    Endoscopic surgical instruments recalled for unvalidated drying process

    Aizu Olympus is recalling cystonephroscope endoscopes due to an unvalidated air drying process that allowed water to remain in instrument channels. Retained moisture poses a risk of contamination and infection.

    Product
    Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0378-2024·2023-11-29

    Biomet Compress Device Anchor Plugs recalled for metal burrs in transverse holes

    Biomet is recalling Compress Device Short Anchor Plugs due to potential metal burrs in the transverse holes that may prevent proper placement of drill pins during surgery, potentially extending surgical time.

    Product
    Compress Device Short Anchor Plug, 10 MM-Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement Item Number: 178552
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0335-2024·2023-11-29

    Hysteroscope Model HYF-1T recalled due to moisture in device channels

    Aizu Olympus is recalling Hysteroscope Model HYF-1T devices because the channel air drying process after repair was not validated. Some returned units had residual moisture that could allow contamination and infection.

    Product
    Hysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0387-2024·2023-11-29

    Biomet Compress Device Short Anchor Plug recalled for metal burrs

    Biomet is recalling the Compress Device Short Anchor Plug due to metal burrs in transverse holes that may prevent proper insertion of drill or pins, potentially extending surgical time. No injuries have been reported.

    Product
    Compress Device Short Anchor Plug, 26 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178568
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0327-2024·2023-11-29

    Bronchoscopes Recalled Due to Inadequate Air Drying Process

    Aizu Olympus recalls bronchoscopes due to unvalidated air drying process. Some returned units after repair contain water in channels, creating infection risk.

    Product
    Bronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190, BF-MP190F, BF-MP60, BF-P150, BF-P190, BF-P60, BF-Q170, BF-Q190, BF-XP190, BF-XP60, BF-XT160, BF-XT
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0364-2024·2023-11-29

    RescuMed Manual Resuscitators Recalled for CO2 Rebreathing Risk

    SunMed Holdings recalls RescuMed manual resuscitators due to a backwards leak in the manometer allowing CO2 rebreathing. Affected models distributed nationwide across 32 US states.

    Product
    Ventlab, LLC RescuMed manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. RescuMed Adult Resuscitator REF #s: BVM510-F, BVMB510-10, BVMB510-F, BVMB510S-F; 2. RescuMed Infant Resuscitator REF #s: BVMB710 & BVMB710-F; 3. RescuMe
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0329-2024·2023-11-29

    Colonoscopes Recalled for Unvalidated Channel Drying Process and Infection Risk

    Aizu Olympus is recalling certain colonoscope models because the channel air drying process was not validated. A small percentage of repaired units had water remaining in the channel, which could lead to infection risk.

    Product
    Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I, CF-2T160L, CF-H170I, CF-H170L, CF-H180AI, CF-H180AL, CF-H180DI, CF-H180DL, CF-H190I, CF-H190L, CF-
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0395-2024·2023-11-29

    Philips MRI Gradient Coil Recall Due to Fire and Smoke Risk

    Philips is recalling SmartPath to dStream 1.5T MRI systems due to a potential gradient coil failure that could produce smoke or fire. The recall affects 119 units distributed nationwide and internationally.

    Product
    SmartPath to dStream for 1.5T
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0396-2024·2023-11-29

    Boston Scientific EXALT Model D Duodenoscope Recall: Fluid Ingress in Lens

    Boston Scientific is recalling certain EXALT Model D Single-Use Duodenoscopes worldwide due to fluid ingress in the lens, which causes poor image quality. Affected devices should not be used; contact the manufacturer.

    Product
    EXALT Model D Single-Use Duodenoscope: Outer box UPN# M00542420 & Outer box UPN# M0054242CE0.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0388-2024·2023-11-29

    Biomet Compress Device Short Anchor Plug recalled due to metal burrs

    Biomet is recalling Compress Device Short Anchor Plugs (28 MM, Item 178570) due to potential metal burrs in transverse holes that may prevent surgical instruments from passing through, extending surgery duration.

    Product
    Compress Device Short Anchor Plug, 28 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178570
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0334-2024·2023-11-29

    Gastrointestinal videoscopes recalled due to unvalidated channel drying process

    Aizu Olympus is recalling 3,627 gastrointestinal videoscopes because the channel air drying process was not validated. A small percentage of devices returned after repair had wet channels, creating potential for contamination and risk of infection.

    Product
    Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q16
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0339-2024·2023-11-29

    Tracheal Videoscope Recall Due to Unvalidated Drying Process

    Aizu Olympus is recalling 61 tracheal videoscopes (Model LF-V) due to an unvalidated air drying process that left water in channels after repair. Water in the endoscope channels may become contaminated and pose an infection risk.

    Product
    Tracheal videoscope- For airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Number: LF-V.
    Category
    Medical Device
    Distribution
    Distributed nationwide