Ultrasonic Bronchoscopes Recalled Due to Inadequate Drying Validation
Aizu Olympus is recalling 823 ultrasonic bronchoscopes (models BF-UC180F, BF-UC190F) because the channel air drying process was not validated, and some repaired units retained water that could cause contamination and infection.
- Product
- Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.
- Category
- Medical Device
- Distribution
- Distributed nationwide