The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

6451–6475 of 114917

  • HighFDA (Devices)·Z-0717-2026·2025-12-03

    IV set 490508 recalled over backflow into the primary container

    B. Braun Medical is recalling IV administration set catalog number 490508 over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Drugs)·D-0179-2026·2025-12-03

    Saline nasal spray recalled over microbial contamination with Pseudomonas lactis

    Medical Products Laboratories is recalling 41,328 bottles of Walgreens Saline Nasal Spray with Xylitol over microbial contamination identified as Pseudomonas lactis.

    Product
    Walgreens Saline Nasal Spray WITH XYLITOL, 1.5 oz (45mL) bottle, DISTRIBUTED BY: WALGREEN CO., 200 WILMOT RD., DEERFIELD, IL 60015, NDC 0363-3114-01
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0671-2026·2025-12-03

    Vented pump sets recalled over medication backflow and occlusion risk

    B. Braun Medical is recalling 12,168 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 15…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0584-2026·2025-12-03

    Blood lead test kits recalled over mislabeled calibration buttons

    Magellan Diagnostics is recalling 1,774 LeadCare II Blood Lead Test Kits because calibration buttons used in kitting were labelled with the wrong lot number.

    Product
    The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare¿ II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0686-2026·2025-12-03

    Low volume IV sets recalled over backflow and priming occlusion

    B. Braun Medical is recalling 960 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMINISTRATION SET, 138…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0646-2026·2025-12-03

    IV set 457506 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 1,019 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET W/2 CARESITE & 1…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0612-2026·2025-12-03

    Anesthesia set 490444 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 88,200 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15D 2 CARESITE…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0711-2026·2025-12-03

    SafeDAY set 352635 recalled over piggyback backflow and inability to prime

    B. Braun Medical is recalling 200,917 SafeDAY sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    SafeDAY set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 60 DROP W/2 SAFEDAY…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0588-2026·2025-12-03

    Heated massage tables recalled over heat components that can smolder

    Pivotal Health Solutions is recalling 464 Quantum Intersegmental Tables because damage to the heat function components can lead to overheating, smoldering and melting of the vinyl cushion.

    Product
    Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0641-2026·2025-12-03

    IV set 490357 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 98,688 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 10…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0615-2026·2025-12-03

    IV set 490349 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 24,648 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    B.BRAUN utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 60 DROPS/ML,141 in.…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0669-2026·2025-12-03

    High volume IV sets recalled over piggyback backflow and occlusion

    B. Braun Medical is recalling 190,625 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DROP W…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0652-2026·2025-12-03

    IV set 490241 recalled over backflow and inability to prime

    B. Braun Medical is recalling 66,792 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET HF…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0714-2026·2025-12-03

    ULTRASITE set 352049-C recalled over backflow and inability to prime

    B. Braun Medical is recalling 41,400 ULTRASITE sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    ULTRASITE set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 2 ULTRASITE VLV…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0664-2026·2025-12-03

    IV set 490315 recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 11,832 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15 DR/ML,2…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0715-2026·2025-12-03

    Vista set US9923F recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 74,652 Vista Basic sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Vista set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. VISTA BASIC PUMP IV SET W/ 15 DROP/ML…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0599-2026·2025-12-03

    Myriad handpieces recalled because packaging tape contains undisclosed latex

    Stryker is recalling 39,148 NICO Myriad Handpieces because tape used to secure components in the packaging contains latex despite the products being labelled latex-free.

    Product
    NICO Myriad Handpiece Product Number/Product Name: NN-8000 15x10 MHP; NN-8001 15x 13 MHP; NN-8002 15x 25 MHP - MINOP; NN-8003 13x10 MHP; NN-8004 13x13 MHP; NN-8005 15x25 MHP GAAB; NN-8006 11X10 MHP; NN-8007 11x13 MHP; NN-8008 19x28 MHP OI; NN-8010…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0624-2026·2025-12-03

    Caresite set 354204 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 12,225,913 units caresite sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. Primary Gravity IV Set with 2…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0594-2026·2025-12-03

    Diabetes clinic software recalled over incorrect sensor glucose graph data

    Medtronic MiniMed is recalling 12,126 CareLink Clinic installations because a software error displays incorrect data on the 24-hour sensor glucose overview graph.

    Product
    CareLink Clinic, REF: MMT-7350
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0665-2026·2025-12-03

    IV set 490489 recalled over piggyback backflow and inability to prime

    B. Braun Medical is recalling 3,120 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15DR, 92 IN…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0645-2026·2025-12-03

    IV set 490067 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 407,800 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET 15DR 5CARESITE 1…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0598-2026·2025-12-03

    Anesthesia set 490469 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 104,784 units anesthesia iv sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Brand Name: Galaxy System Product Name: Galaxy System Model/Catalog Number: GAL-001 Product Description: The Galaxy SystemTM and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedure.
    Category
    Medical Device
    Distribution
    20 states
  • HighFDA (Food)·H-0207-2026·2025-12-03

    Ice cream bars recalled over undeclared wheat and soy allergens

    Jeni's Splendid Ice Creams is recalling 3,767 cases of Passion Fruit Dreamsicle Ice Cream Bars over undeclared wheat and soy.

    Product
    Jeni's Passion Fruit Dreamsicle Ice Cream Bars, 3 FL OZ (89 ML)/9 FL OZ (266 ML), packaged 3 bars per retail carton with Item UPC: 850060644122, packed 8 retail cartons per case with Case UPC: 10850060644129
    Category
    Food
    Distribution
    10 states
  • HighFDA (Devices)·Z-0706-2026·2025-12-03

    Outlook infusion sets recalled over backflow and inability to prime

    B. Braun Medical is recalling 6,672 Outlook sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK PUMP SET,3 CARESITE LADS,115…
    Category
    Medical Device
    Distribution
    0 states