I.T.S. hand locking plates heat more in MRI than instructions state
I.T.S. GmbH is recalling the HLS Hand Locking Plates System because updated MRI safety testing shows higher RF-induced temperature increases than the Instructions for Use previously reflected.
- Product
- HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11, 32151-12, 32151-13, 32151-14, 32151-15, 32151-16, 32151-18, 32151-20. 2. HLS Plate, 2 T-Shape
- Category
- Medical Device
- Distribution
- Distributed nationwide