The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

1151–1175 of 27681

  • HighFDA (Devices)·Z-1779-2026·2026-04-15

    Blue Rhino G2-Multi tracheostomy introducer sets carry overstated expiration dates

    Cook Medical is recalling Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Sets because affected lots were labeled with expiration dates that exceed the true shelf life. 916 units are affected.

    Product
    COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-HC-G-EU, Order Number G57685; Reference Part Number C-PTIS-100-HC-G-NA, Order Number G57682; Reference Part Number C-PTIS-100-UNL-HC-G-EU, Order Number G57687…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1791-2026·2026-04-15

    Cook Medical Wayne pneumothorax sets carry overstated expiration dates

    Cook Medical is recalling Wayne Pneumothorax Sets because affected lots were labeled with expiration dates that exceed the product's true shelf life. 104 units are affected.

    Product
    COOK MEDICAL Wayne Pneumothorax Set: Reference Part Number C-UTPT-1020-WAYNE-IMH, Order Number G56532; Reference Part Number C-UTPT-1400-WAYNE-112497-IMH, Order Number G56535.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1795-2026·2026-04-15

    Preeclampsia growth factor assay recalled over out-of-range quality control values

    B.R.A.H.M.S PlGF Plus KRYPTOR in vitro diagnostic tests are being recalled after customers observed out-of-range values at the lowest quality control level, which can contribute to falsely elevated results.

    Product
    B¿R¿A¿H¿M¿S KRYPTOR compact REACT, REF: 89986, used with B¿R¿A¿H¿M¿S PlGF Plus KRYPTOR, REF: 859075N, and run on the B¿R¿A¿H¿M¿S KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor
    Category
    Medical Device
    Distribution
    15 states
  • HighFDA (Devices)·Z-1804-2026·2026-04-15

    Stryker tourniquet cuffs may detach from the bladder welding connection

    Stryker Sustainability Solutions is recalling Color Cuff tourniquet cuffs because the cuff can detach from the bladder at the welding connection, leaving it unable to maintain pressure.

    Product
    Stryker Sustainability Solution Color Cuff 24" (Yellow) Quick Connect Dual Port Single Bladder REF 5921-024-235 Stryker Sustainability Solution Color Cuff 18" (Red), Quick Connect Dual Port Single Bladder REF 5921-018-135 Stryker Sustainability Solution Color Cuff 18"…
    Category
    Medical Device
    Distribution
    41 states
  • HighFDA (Devices)·Z-1769-2026·2026-04-15

    Tornier shoulder system parts mislabeled as compatible with the HRS Max

    Tornier HRS Max tuberosity bodies and a trial component were incorrectly labeled as HRS Max products; they are compatible only with the Tornier HRS system. 233 units are affected.

    Product
    Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1778-2026·2026-04-15

    Cook Medical arterial pressure monitoring trays carry overstated expiration dates

    Cook Medical is recalling Arterial Pressure Monitoring Trays because affected lots were labeled with expiration dates that exceed the product's true shelf life. 20 units are affected.

    Product
    COOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1786-2026·2026-04-15

    Cook Medical multipurpose drainage catheters carry overstated expiration dates

    Cook Medical is recalling a Multipurpose Drainage Catheter because affected lots were labeled with expiration dates that exceed the product's true shelf life. 20 units are affected.

    Product
    COOK MEDICAL Multipurpose Drainage Catheter: Reference Part Number ULT8.5-38-25-P-CLMS-8-RH-FS, Order Number G29775
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1773-2026·2026-04-15

    Xtant pedicle screws mislabeled with an incorrect shorter screw length

    Xtant Medical is recalling Xpress Pedicle Screw System dual lead screws from one lot that were mislabeled with a part number identifying them as shorter than they physically are.

    Product
    XTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-1781-2026·2026-04-15

    Cook Medical tracheostomy introducer sets labeled with overstated expiration dates

    Cook Medical is recalling Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Sets because affected lots were labeled with expiration dates that exceed the true shelf life. 484 units are affected.

    Product
    COOK MEDICAL Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-FLEX-HC, Order Number G32092; Reference Part Number C-PTIS-100-HC, Order Number G12115.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1787-2026·2026-04-15

    Cook Medical nitinol stone extractors labeled with overstated expiration dates

    Cook Medical is recalling NCompass Nitinol Stone Extractors because affected lots were labeled with expiration dates that exceed the product's true shelf life. 70 units are affected.

    Product
    COOK MEDICAL NCompass¿ Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250; Reference Part Number C-NTSE-2.4-115-NCT4, Order Number G36251.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1776-2026·2026-04-15

    Cook Medical micro wire guides labeled with overstated expiration dates

    Cook Medical is recalling Approach CTO-12 Micro Wire Guides because affected lots were labeled with expiration dates that exceed the product's true shelf life. 49 units are affected.

    Product
    COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-1647-2026·2026-04-08

    Trilogy Evo O2 obstruction alarm may take up to four breaths

    Philips Respironics is recalling the Trilogy Evo O2 ventilator because in some situations the Obstruction Alarm does not trigger within the timeframe required by the relevant standards.

    Product
    Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1649-2026·2026-04-08

    Trilogy Evo Universal obstruction alarm may take up to four breaths

    Philips Respironics is recalling the Trilogy Evo Universal ventilator because in some situations the Obstruction Alarm does not trigger within the timeframe required by the relevant standards.

    Product
    Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1646-2026·2026-04-08

    Trilogy Evo obstruction alarm may take up to four breaths

    Philips Respironics is recalling the Trilogy Evo ventilator because in some situations the Obstruction Alarm does not trigger within the timeframe required by the relevant standards.

    Product
    Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1637-2026·2026-04-08

    Volara ventilator adapter recalled as Class I over desaturation and barotrauma

    Baxter is issuing an urgent correction for the Blue Ventilator Adapter Module used with the Volara system because home users may be unaware of possible oxygen desaturation or lung injury from overexpansion.

    Product
    Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937. Accessory to Baxter Volara Respiratory Therapy system.
    Category
    Medical Device
    Distribution
    1 state
  • SevereFDA (Devices)·Z-1591-2026·2026-04-08

    Impella controller purge blocked alarm may be delayed on older pumps

    Abiomed is recalling Automated Impella Controllers because the Purge System Blocked alarm display could be delayed when using first generation Impella 5.5 pumps.

    Product
    Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK…
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-1648-2026·2026-04-08

    Trilogy EV300 obstruction alarm may take up to four breaths

    Philips Respironics is recalling the Trilogy EV300 ventilator because in some situations the Obstruction Alarm does not trigger within the timeframe required by the relevant standards.

    Product
    Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1589-2026·2026-04-08

    Impella purge cassettes recalled as Class I over purge leak risk

    Purge Cassettes sold within Impella Pump Sets and individually packaged are being recalled because of an increased risk of purge leaks with Generation 1 purge cassettes.

    Product
    Purge Cassette, sold within Impella Pump Sets and Individually Packaged. Labeled and distributed in the below configurations and corresponding product codes. 1. Purge Cassette, 5 Pack; Product Code: 0043-0003. 2. Impella RP US Pump Set; Product Code: 004334. 3. Purge…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1679-2026·2026-04-08

    Philips Azurion 5 M12 X-ray may not start from foot switch

    Philips is recalling the Azurion 5 M12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 5 M12; System Code: (1)722227, (2)722231;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1702-2026·2026-04-08

    VOLT locking screw taps swapped between lots carry incorrect thread

    The 2.4 VOLT Tap Locking Screw Tap is being recalled after a full lot swap with the Cortical Screw Tap left the bone taps with an incorrect thread.

    Product
    2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1691-2026·2026-04-08

    Medline ENFit G-tube connector kits may leak from dimensional defect

    Medline is recalling kits containing ENFit G-tube connectors because the connectors were not made to required dimensional specifications and may not form a proper seal.

    Product
    Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1708-2026·2026-04-08

    Medline endoscopy kits contain recalled Olympus biopsy valves shedding fragments

    Medline is recalling BAL and ENDO kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1747-2026·2026-04-08

    MiniMed 620G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 620G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 620G Insulin Pump (MMT-1750)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1668-2026·2026-04-08

    Philips Allura Xper FD20/10 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD20/10 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/10; System Code: 722029;
    Category
    Medical Device
    Distribution
    Distributed nationwide