The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

35251–35275 of 40325

  • LowFDA (Devices)·Z-2129-2014·2014-08-06

    Septodont Recalls Compolite Bond SE Due to Incorrect Packaging

    Septodont is recalling 24 units of Compolite Bond SE dental adhesive because the material was incorrectly packaged. The product was distributed to wholesalers in Pennsylvania and New York.

    Product
    Darby Dental Supply Compolite Bond SE Self-Etch/Light-Curing. This product is used to adhere restorative material to the tooth during restorations. Package contains 100 0.125 mL unit dose tips, 100 applicator brushes, and a product insert.
    Category
    Medical Device
    Distribution
    2 states
  • LowFDA (Devices)·Z-2133-2014·2014-08-06

    Siemens Dimension Integrated Chemistry Systems Recalled for Photometer Misassembly

    Siemens is recalling 15 Dimension Integrated Chemistry Systems due to a photometer misassembly issue that may shorten the life of the source lamp.

    Product
    Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.
    Category
    Medical Device
    Distribution
    9 states
  • LowFDA (Devices)·Z-2128-2014·2014-08-06

    Septodont Recalls Self-Etch Bond Adhesive Due to Packaging Error

    Septodont is recalling Self-Etch bond adhesive because the material was incorrectly packaged. The recall involves 60 units distributed to two wholesalers in Pennsylvania and New York.

    Product
    Henry Schein Natural Elegance SE Bond Self-Etch Light-Cure Adhesive contains 50 0.125 mL unit dose tips, 50 applicator brushes, and a product insert.
    Category
    Medical Device
    Distribution
    2 states
  • SevereFDA (Devices)·Z-2068-2014·2014-07-30

    GE Healthcare Recalls DateX-Ohmeda S/5 Airway Modules Due to CO2 Measurement Failure

    GE Healthcare is recalling DateX-Ohmeda S/5 Single-Width Airway Modules and E-miniC accessories because a CO2 detector failure may cause incorrect low readings, potentially impairing clinical decision-making for ventilated patients.

    Product
    DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES. DATEX-OHMEDA S/5TM FM WITH L-FICU04 ANDL-FICUO4A SOFTWARE AND N-FCREC MODULE **(OUS only). E-miniC and accessories is indicated for monitoring C02 and respiration rate of all hospital patients.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2080-2014·2014-07-30

    Stryker Recalls Disposable StrykeFlow Tips Due to Sterility Concerns

    Stryker Endoscopy is recalling 496 boxes of Disposable StrykeFlow Tips because a deformed blister pouch could compromise the product's sterility.

    Product
    Disposable StrykeFlow Tip, Stryker Disposable Suction/I irrigator Tip; Sterile, Rx only, Latex Free. Stryker Suction Irrigation system, used to provide a clean and clear operative site through the evacuation and aspiration of irrigant solution. for endoscopic use.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2021-2014·2014-07-30

    Blood Glucose Test Strips Recalled for Inappropriate Labeling and Marketing

    Diabetic Supply of Suncoast, Inc. is recalling Advocate Redi-Code+ blood glucose test strips because they were inappropriately labeled and marketed.

    Product
    ADVOCATE Redi-Code+ BMB-BA006A Blood Glucose test strips Product Usage: The ADVOCATE¿ Redi-Code+ BMB-BA006A Blood Glucose test strips are used only with the ADVOCATE¿ Redi-Code+ BMB-EA004S meter to quantitatively measure glucose in fresh capillary whole blood samples drawn…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2116-2014·2014-07-30

    Pentax Ultrasound Gastroscopes Recalled Due to Aspiration Needle Derailment Risk

    Pentax Medical is recalling 531 ultrasound gastroscopes because aspiration needles may derail from the accessory elevator, posing a risk of patient injury.

    Product
    EG-3630U Ultrasound Video Gastroscope EG-3830UT Ultrasound Video Gastroscope EG-3870UTK Ultrasound Video Gastroscope FG-36UX Ultrasound Fiber Gastroscope Used to provide optical and sonographic visualization of and therapeutic access to the Upper Gastrointestinal Tract.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2115-2014·2014-07-30

    Heartsine AED Electrode Pouches Recalled for Difficulty Opening

    Heartsine is recalling Pad-Pak and Pedi-Pak electrode accessories for the Samaritan PAD 300P and 350P because some sealed foil pouches are difficult to open, potentially delaying therapy.

    Product
    Pad-Pak and Pedi-Pak accessories for the HeartSine Samaritan PAD 300P and 350P device (SAM 300P and 350P). Automated External Defibrillator electrode supplies. The Pedi-Pak is specifically designed for use with the HeartSine Samaritan PAD 300P and 350P device. It contains…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2100-2014·2014-07-30

    Terumo Recalls Sarns Two-stage Venous Return Cannulae Due to Particulate

    Terumo CVS is recalling Sarns Two-stage Venous Return Cannulae because in-process inspections found loose fiber particulate exceeding specifications.

    Product
    Sarns Two-stage Venous Return Cannulae 32/40 & 36/51 Fr with 1/2" & 3/8" flare or connector Product Usage: The Venous Return Cannula is indicated for venous drainage during cardiopulmonary bypass surgery for dual cannulation of the superior and inferior vena cava.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2109-2014·2014-07-30

    Philips QLAB 10 Core Module Software Recall for Measurement Defect

    Philips Ultrasound is recalling QLAB 10 Core Module software due to a defect that may use incorrect measurement values under specific conditions.

    Product
    QLAB 10 Core Module PN 453561704771. QLAB Quantification Software is a software application package designed to view and quantify image data acquired on Philips Medical Systems ultrasound products.
    Category
    Medical Device
    Distribution
    10 states
  • HighFDA (Devices)·Z-2114-2014·2014-07-30

    GE Healthcare Recalls CARESCAPE Monitor B650 Due to Sound Loss

    GE Healthcare is recalling CARESCAPE Monitor B650 units because they may lose alarm sounds when an external display is connected via a USB extender with an active hub.

    Product
    GE Healthcare, CARESCAPE Monitor B650. A multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2108-2014·2014-07-30

    Teleflex Recalls Hudson RCI Breathing Circuits Due to Temperature Probe Defect

    Teleflex Medical is recalling Hudson RCI breathing circuits because the temperature probe does not properly connect to the temperature port.

    Product
    Hudson RCI, Breathing Circuit, Adult Single Limb (W/Connector, Adaptor, Y Piece), Rx Only, Single Patient Use. The Hudson Heated Wire Ventilator Circuit is intended as a conduit for respiratory gas between a patient and a ventilator, and includes heated wires for use with a…
    Category
    Medical Device
    Distribution
    16 states
  • HighFDA (Devices)·Z-2089-2014·2014-07-30

    Perouse Flamingo Inflation Devices Recalled for Potential Water-Tightness Defect

    Perouse Medical is voluntarily recalling specific batches of Flamingo Inflation Devices due to a potential water-tightness defect that could allow air entry during cardiovascular procedures.

    Product
    Perouse Flamingo, Model 0218TA and 0218TB intended for use during cardiovascular procedure to create, maintain and monitor pressure in the balloon catheter.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2082-2014·2014-07-30

    DePuy RECLAIM Taper Protector Sleeve Recalled for Removal Difficulty

    DePuy Orthopaedics is recalling 814 RECLAIM Taper Protector Sleeves because the product can be difficult to remove from the Distal Stem.

    Product
    RECLAIM Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during proximal preparation.
    Category
    Medical Device
    Distribution
    49 states
  • HighFDA (Devices)·Z-2085-2014·2014-07-30

    GE Healthcare CARESCAPE Patient Monitors Recalled for Data Loss

    GE Healthcare is recalling CARESCAPE B850, B650, and B450 patient monitors because the Patient Data Module may fail to communicate, causing partial loss of monitoring parameters.

    Product
    GE Healthcare, CARESCAPE Monitor B850, B650 or B450. Patient Monitor.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2111-2014·2014-07-30

    Siemens syngo Imaging XS Software Recall for Potential Miscalculation

    Siemens is recalling syngo Imaging XS software due to potential miscalculation of grey scale values in specific image processing functions.

    Product
    Siemens syngo Imaging XS, version VA70A or higher Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2091-2014·2014-07-30

    Invacare Recalls M41 Wheelchair Controllers Due to Programming Error

    Invacare is recalling 819 M41 powered wheelchair controllers that may cause momentary loss of control due to non-standard factory programming.

    Product
    MK5 NX Controller--Wheelchair controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2096-2014·2014-07-30

    Terumo Recalls Sarns Venous Return Cannulae Due to Loose Fiber Particulate

    Terumo CVS is recalling Sarns Malleable Dual-stage Venous Return Cannulae because in-process inspections found loose fiber particulate exceeding specifications.

    Product
    Sarns Malleable Dual-stage Venous Return Cannulae 28/38, 32/40 & 34/46 Fr with 1/2" connector with or without Xcoating Product Usage: The Venous Return Cannula is indicated for venous drainage during cardiopulmonary bypass surgery for dual cannulation of the superior and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2092-2014·2014-07-30

    Alphatec Spine Recalls Solanas 2.1mm Fixed Depth Drill

    Alphatec Spine is recalling 14 units of the Solanas 2.1mm Fixed Depth Drill because the golden stop slid from its fixed position during use.

    Product
    Solanas 2.1mm FIXED DEPTH DRILL, 10mm, Part No. 63928-10. Component of the SOLANAS III Posterior Stabilization System, a spinal fixation system intended to improve stability of the cervical and thoracolumbar area of the spine (C1-T3).
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2099-2014·2014-07-30

    Terumo Recalls Sarns Soft-Flow Aortic Cannulae Due to Fiber Particulate

    Terumo CVS is recalling Sarns Soft-Flow Aortic Cannulae because in-process inspections found loose fiber particulate exceeding specifications.

    Product
    Sarns Soft-Flow Extended Aortic Cannulae 7.0mm & 8.0mm with or without luer & Xcoating, Product Usage: The Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2098-2014·2014-07-30

    Terumo Recalls Sarns Soft-arc Aortic Cannula Due to Loose Fiber Particulate

    Terumo CVS is recalling 20 units of Sarns Soft-arc Aortic Cannulas because in-process inspections found loose fiber particulate exceeding specifications.

    Product
    Sarns Soft-arc Aortic Cannula 8.0mm curved tip, suture flange, 3/8" flare Product Usage: The Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2079-2014·2014-07-30

    Siemens RAPIDPoint 500 Lactate Cartridges Recalled for Electrolyte Error Codes

    Siemens Healthcare Diagnostics Inc is recalling RAPIDPoint 500 Lactate Cartridges because they may display error codes for electrolyte parameters.

    Product
    RAPIDPoint 500 Measurement Clinical Chemistry System RAPIDPoint 500 Measurement Cartridge Lactate (750 Tests) Part number: 10491449; RAPIDPoint 500 Measurement Cartridge Lactate (400 Tests) Part Number: 10491448 RAPIDPoint 500 Measurement Cartridge Lactate (250 Tests) Part…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2094-2014·2014-07-30

    Terumo Recalls Sarns D4 Aortic Arch Cannulae Due to Particulate

    Terumo CVS is recalling Sarns D4 Aortic Arch Cannulae because in-process inspections found loose fiber particulate exceeding specifications.

    Product
    Sarns D4 Aortic Arch Cannulae 6.0mm, 7mm & 8mm angled and straight tip with or without side holes and/or luer Product Usage: The Sarns Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2081-2014·2014-07-30

    AGFA DX-D 100 X-Ray System Recalled for Touch Screen Malfunction

    AGFA Corp. is recalling 216 DX-D 100 digital radiography systems because liquid contact with the touch screen may cause the device to incorrectly recognize user input and alter settings.

    Product
    AGFA Digital Radiography X-Ray system DX-D 100 Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton including skull…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2084-2014·2014-07-30

    DePuy RECLAIM Taper Protector Sleeve Recalled for Removal Difficulty

    DePuy Orthopaedics is recalling RECLAIM Taper Protector Sleeves because they can be difficult to remove from the Distal Stem.

    Product
    RECLAIM DISTAL TAPERED Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during proximal preparation.
    Category
    Medical Device
    Distribution
    49 states