The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

29076–29100 of 40325

  • HighFDA (Devices)·Z-2725-2016·2016-09-14

    Hospira Recalls Plum A+3 Infusion Pumps Due to Undersized Diaphragm

    Hospira is recalling 357,778 Plum A+3 Infusion Pumps because an undersized fluid shield diaphragm may cause alarms or therapy interruption.

    Product
    Plum A+3 Infusion Pump. List Number 12391. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self-contained, microprocessor…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2769-2016·2016-09-14

    Innovision Innocor Cardiopulmonary Testing System Recalled for Electric Shock Risk

    Innovision A/S is recalling eight Innocor Cardiopulmonary Exercise and Lung Function Testing Systems to replace USB cables that lack galvanic isolation, preventing potential electric shock.

    Product
    Innocor Cardiopulmonary Exercise and Lung Function Testing System Non-invasive diagnostic device that measures or calculates a number of hemodynamic and metabolic parameters.
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-2765-2016·2016-09-14

    Hologic Recalls SureSight Obturator Introducer Systems Due to Tip Breakage

    Hologic is recalling 1,440 SureSight Obturator Introducer Localization Systems because the tip may break at the glue joint.

    Product
    SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. ILS-0914-20-OB (9G Introducer Localization Set (OB Model) for ATEC0914-20MR Handpiece)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2737-2016·2016-09-14

    Philips BrightView XCT Upgrade Recalled for Potential Patient Contact

    Philips is recalling three BrightView XCT Upgrade units due to a software defect that may cause the detector to contact patients during specific scans.

    Product
    BrightView XCT Upgrade Model 882454
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2729-2016·2016-09-14

    Hospira Recalls Plum A+3 Infusion Pumps Due to Undersized Diaphragm

    Hospira Inc. is recalling 357,778 Plum A+3 Infusion Pumps because an undersized fluid shield diaphragm may cause alarms and interrupt therapy.

    Product
    Plum A+3 Infusion Pump. List Number 12618. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self-contained, microprocessor…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2734-2016·2016-09-14

    Omega Medical Imaging Recalls Elevating Monitor Suspension Systems

    Omega Medical Imaging is recalling 26 Elevating Monitor Suspension Systems because the suspension may separate from the lifting column bracket.

    Product
    2 over 2 Elevating Monitor Suspension - Omega Medical Imaging, LLC Elevating Monitor Suspension Systems.
    Category
    Medical Device
    Distribution
    14 states
  • HighFDA (Devices)·Z-2738-2016·2016-09-14

    Terumo Recalls Pinnacle Introducer Sheaths Due to Valve Leakage

    Terumo Medical Corporation is recalling Pinnacle Introducer Sheaths due to a manufacturing issue causing valve leakage, which poses a risk of incremental blood loss.

    Product
    PINNACLE¿ Introducer Sheath PINNACLE PRECISION ACCESS SYSTEM¿ Sheath PINNACLE¿ TIF TIP" Introducer Sheath PINNACLE¿ R/O II Radiopaque Marker Introducer Sheath
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2762-2016·2016-09-14

    Riverpoint Medical Recalls Webglo Polypropylene Surgical Sutures Due to Seal Issues

    Riverpoint Medical is recalling specific lots of Webglo polypropylene surgical sutures because the seals may open during distribution or packaging.

    Product
    Polypropylene Surgical Sutures, brand Webglo, for vet use. Product is sterile. Part number: GP8665 Product Usage: For use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurological surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2763-2016·2016-09-14

    Blackrock NeuroPort Biopotential Signal Processing System Recall for Labeling Error

    Blackrock Microsystems is recalling 81 Neural Signal Amplifiers because the device and patient cable are incorrectly labeled as Type CF Applied Parts.

    Product
    Neural Signal Amplifier. Part Numbers: 4208, 5703, 5749, 5748, 5747. Product Usage: The Blackrock NeuroPort Biopotential Signal Processing System supports recording, processing and display of biopotential signals from user supplied electrodes. Biopotential signals include…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2764-2016·2016-09-14

    Blackrock NeuroPort Patient Cables Recalled for Incorrect Labeling

    Blackrock Microsystems is voluntarily recalling 120 Patient Cables because they are incorrectly labeled as Type CF Applied Parts.

    Product
    Patient Cable. Part Numbers: 4460, 6735. Sterile within Chipboard Boxes within corrugated cartons. Product Usage: The Blackrock NeuroPort Biopotential Signal Processing System supports recording, processing and display of biopotential signals from user supplied electrodes.…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2742-2016·2016-09-14

    Siemens ADVIA Centaur PSA Assay Recall for Clinical Interpretation Clarification

    Siemens Healthcare Diagnostics is recalling ADVIA Centaur PSA Assay kits to emphasize that PSA values must be interpreted according to current clinical guidelines for biochemical recurrence.

    Product
    ADVIA Centaur¿ PSA Assay; Cat No. 06574155/SMN 10310292 (100 test)/02676506/SMN 10310293 (500 test)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2756-2016·2016-09-14

    Sorin Aortic Arch Cannula Recalled for Improper Tip Angle

    Sorin Group USA is recalling Aortic Arch Cannulas because the tip angle is less than the required 45 or 90 degrees.

    Product
    Aortic Arch Cannula and Coronary Artery Perfusion Cannula with Ballon (part number NA-55X6). The Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced…
    Category
    Medical Device
    Distribution
    3 states
  • ModerateFDA (Devices)·Z-2747-2016·2016-09-14

    Mentor Breast Tissue Expanders Recalled Due to Product Mix-Up

    Mentor Texas, LP. is recalling 65 units of MENTOR CPX4 breast tissue expanders because a 350cc unit was found in a carton intended for 450cc units.

    Product
    MENTOR 2" Tall Height TE With Sutures 350cc PN: 354-9312 The MENTOR CPX4 and CPX4 with Suture Tabs Breast Tissue Expanders are used for breast reconstruction following mastectomy and are intended for temporary subcutaneous or submuscular implantation and are not intended for…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2743-2016·2016-09-14

    Integra Recalls Achillon Suture System Due to Sterility Concerns

    Integra LifeSciences is recalling the Achillon Suture System because internal studies suggest the secondary package seal may not maintain intended sterility.

    Product
    Achillon Minimally Invasive Device for Achilles Tendon Rupture Treatment The Achillon Suture System is a single use, sterile device used to treat acute Achilles tendon ruptures. The Achillon Suture System is a mini-invasive procedure indicated to teat acute Achilles tendon…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2748-2016·2016-09-14

    Mentor CPX4 Breast Tissue Expanders Recalled for Product Mix-Up

    Mentor Texas, LP. is recalling 70 units of CPX4 breast tissue expanders because a 350cc product was found in a carton labeled for the 450cc version.

    Product
    MENTOR 2" Tall Height TE With Sutures 450cc PN: 354-9313 The MENTOR CPX4 and CPX4 with Suture Tabs Breast Tissue Expanders are used for breast reconstruction following mastectomy and are intended for temporary subcutaneous or submuscular implantation and are not intended for…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2744-2016·2016-09-14

    Ecolab Recalls Microtek Angiography Drapes Due to Packaging Defects

    Ecolab is recalling specific Microtek angiography drapes because their sterile packaging may contain small holes in the film, potentially compromising sterility.

    Product
    Microtek Franz Angiography Drape with Radiation Shield, Item # RBA202, and Microtek Angiography Drape with Radiation Shield and Pouches, Item # RBA203. Sterile EO. Microtek Medical, Inc., Columbus, MS 39702.
    Category
    Medical Device
    Distribution
    4 states
  • ModerateFDA (Devices)·Z-2752-2016·2016-09-14

    Zimmer Biomet Recalls Coonrad/Morrey Elbow Replacement Kits Due to Missing Parts

    Zimmer Biomet is recalling a specific lot of Coonrad/Morrey Elbow Pin/Bushing Replacement Kits because they are missing a humeral bushing, which could delay surgery.

    Product
    Coonrad/ Morrey Elbow Pin/ Bushing Replacement size Kit Extra Small. This product is sold sterile. The parts are placed into an inner cavity and are sealed using a nitrogen process. The inner cavity is then placed into an outer cavity and heat sealed with outer Tyvek lid on the…
    Category
    Medical Device
    Distribution
    9 states
  • ModerateFDA (Devices)·Z-2767-2016·2016-09-14

    Novadaq Recalls SK6000 PAQ Sterile Drape Pouches Due to Pinhole Defect

    Novadaq is recalling 20 units of SK6000 PAQ sterile drape pouches because pinholes were found in the film material during shipping validation testing.

    Product
    Sterile Drape, SK6000 PAQ drape pouch (for use with the SPY Fluorescent System (SPY 2000/2001)) The SPY Fluorescent Imaging System is intended to provide fluorescent images for the visual assessment of blood flow in vessels and related tissue perfusion during cardiovascular…
    Category
    Medical Device
    Distribution
    3 states
  • ModerateFDA (Devices)·Z-2755-2016·2016-09-14

    Aspen Surgical Recalls Misidentified Stainless Steel Surgical Blades

    Aspen Surgical Puerto Rico Corp. is recalling 6,000 stainless steel surgical blades that were incorrectly labeled as size 12 instead of 12B.

    Product
    Intermediate carton label reads in part as: "Bard-Parker***Stainless Steel Surgical Blades***#12B***QTY 150 (3 x 50)***REF 371712***LOT 0088464." Intended for tissue separation and other procedures that require a sharp surgical blade to puncture or cut.
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-2754-2016·2016-09-14

    Anthogyr Recalls Axiom 2.8 Dental Implants Due to Design Validation Failure

    Anthogyr is recalling 10 units of Axiom 2.8 dental implants because an FDA inspection revealed the firm failed to validate the design and process of the devices.

    Product
    Axiom 2.8 implants Sterile The Axiom REG implants are used as artificial root structures for replacement of missing teeth. They can be used for stabilization of removable prostheses or fixation of single tooth restorations or partial dentures. There are different diameters…
    Category
    Medical Device
    Distribution
    3 states
  • ModerateFDA (Devices)·Z-2753-2016·2016-09-14

    Anthogyr Axiom REG Dental Implants Recalled for Design Validation Failure

    Anthogyr is recalling 111 units of Axiom REG dental implants because an FDA inspection found the firm failed to validate the design and process of the devices.

    Product
    Axiom REG implants Sterile The Axiom REG implants are used as artificial root structures for replacement of missing teeth. They can be used for stabilization of removable prostheses or fixation of single tooth restorations or partial dentures. There are different diameters…
    Category
    Medical Device
    Distribution
    3 states
  • ModerateFDA (Devices)·Z-2759-2016·2016-09-14

    Teleflex Horizon Manual-Load Ligating Clip Applier Recalled for Incorrect Etching

    Teleflex Medical is recalling 49 units of Horizon Manual-Load Ligating Clip Appliers due to incorrect etching on the device.

    Product
    WECK¿ Horizon" Manual-Load Ligating Clip Applier, For Medium-Large Size Horizon Ligation Clips Curved Jaw, 11 (28cm), Rx Only, REF 337111, Product Usage: Weck Closure Systems applier has been designed and calibrated for use with specific Horizon, Hemoclip Plus Hemoclip…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2720-2016·2016-09-14

    ReliaMed Alginate Dressings Recalled for Misplaced Instructions

    Advanced Medical Solutions is recalling ReliaMed Alginate/CMC dressings because instructions for use were misplaced in the packaging.

    Product
    ReliaMed¿ Alginate/CMC. Each dressing is packed in a pouch (primary packaging), each pouch is packed in a carton (secondary packaging) holding multiple (depending on stock keeping unit) pouches and each carton contains one (1) Instructions For Use (IFU) leaflet.
    Category
    Medical Device
    Distribution
    1 state
  • ModerateFDA (Devices)·Z-2757-2016·2016-09-14

    Sorin Coronary Artery Perfusion Cannula Recalled for Tip Angle Defect

    Sorin Group USA is recalling 1,205 units of Coronary Artery Perfusion Cannulas with Balloons because the tip angle is less than the required 45 or 90 degrees.

    Product
    Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided for with a vinyl female luer connector bonded proximally and a vinyl cuff bonded distally. The…
    Category
    Medical Device
    Distribution
    3 states
  • ModerateFDA (Devices)·Z-2739-2016·2016-09-14

    GE Healthcare Recalls Centricity Enterprise Archive Software Due to Data Loss Risk

    GE Healthcare is recalling Centricity Enterprise Archive software version 3.0.x due to a potential for data loss when the cache disk is nearly full.

    Product
    Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. Qualified system administrators install, monitor and maintain the system. DICOM devices (e.g. modalities, workstations) communicate with the archive using the DICOM protocol…
    Category
    Medical Device
    Distribution
    Distributed nationwide