Agency classClass IIOur severity:HighFDA (Devices)·Z-3362-2018·Announced 2018-10-10

B. Braun Medical Recalls ACCU-BLOC PERIFIX Kits Due to Leakage

B. Braun Medical is recalling CE18TK ACCU-BLOC PERIFIX Kits because the catheter connectors may not stay closed, causing leakage or disconnection.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device with a functional defect (leakage/disconnection) that poses a risk of harm to patients, but the source text does not report any specific injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.

Plain-English summary

B. Braun Medical, Inc. is recalling the CE18TK ACCU-BLOC PERIFIX KIT, Material Number 332220. This connection device is used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for the delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.

The recall was initiated because the catheter connectors may not stay closed during use. In some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.

Approximately 63,980 units were distributed nationwide in the United States and in Australia, Canada, Guatemala, and Kuwait. The affected batch numbers are 61595170, 61595613, 61596536, 61598356, 61597084, 61600180, 61617168, 61616702, 61617169, 61616703, 61614599, 61616701, 61616700, 61614596, 61614598, 61619583, 61620763, 61622163, and 61621828. Healthcare providers should stop using these devices and contact B. Braun Medical for instructions on how to return the product.

The recalled product

Product
CE18TK ACCU-BLOC PERIFIX KIT, Material Number 332220 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with…
Affected units
63,980

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (19)

  • Batch: 61595170
  • 61595613
  • 61596536
  • 61598356
  • 61597084
  • 61600180
  • 61617168
  • 61616702
  • 61617169
  • 61616703
  • 61614599
  • 61616701
  • 61616700
  • 61614596
  • 61614598
  • 61619583
  • 61620763
  • 61622163
  • 61621828

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 93 recalled by B. Braun Medical, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 93 products in this recall →