Ortho-Clinical VITROS Chemistry CREA Slides Recalled for Measurement Accuracy Issues
Ortho-Clinical Diagnostics is recalling VITROS Chemistry CREA Slides because they failed to meet accuracy standards for low-end creatinine measurements in serum and urine.
The agency’s classification
FDA (Devices): Class III
Class III is a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 1/5 (Low).
Cosmetic or documentation issue. No expected health impact, but the manufacturer is correcting the affected products.
Our severity reasoning: The recall is classified as FDA Class III, which by definition is unlikely to cause adverse health consequences. The issue is a technical accuracy defect in the measurement range rather than a direct physical hazard.
Plain-English summary
Ortho-Clinical Diagnostics is recalling VITROS Chemistry Products CREA Slides (Product Code 680 2584) due to performance issues with the assay's measurement range. The product is used for in vitro diagnostic testing to quantitatively measure creatinine concentration in serum, plasma, and urine.
The recall was initiated because the slides did not meet the current claims for Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantification (LoQ) for serum and urine samples. This failure affects the low end of the measuring range. The original low-end measuring range was 0.05 mg/dL for serum and 1.2 mg/dL for urine, but testing results indicate the revised low-end range is 0.15 mg/dL for serum and 3.2 mg/dL for urine.
The affected units were distributed worldwide, including the United States, Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, and the United Kingdom. The recall covers 627,543 units across generations 7, 10, 12, 13, 15, 16, and 18. Users should contact Ortho-Clinical Diagnostics for further instructions regarding the return or replacement of these slides.
The recalled product
- Product
- VITROS Chemistry Products CREA Slides, Product Code 680 2584 Product Usage: VITROS CREA Slides (Product Code 680 2584) - For in vitro diagnostic use only. VITROS Chemistry Products CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and…
- Manufacturer
- Ortho-Clinical Diagnostics
- Hazard
- Affected units
- 627,543
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (7)
- GEN 7
- GEN 10
- GEN 12
- GEN 13
- GEN 15
- GEN 16
- GEN 18
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Ortho-Clinical Diagnostics
See all →- HighClinical Laboratory Quality Control Device Recalled for Diluent Assembly Error
FDA (Devices) · 2024-04-03
- HighLaboratory performance verifier recalled for mismatched diluent assembly
FDA (Devices) · 2024-04-03
- HighVITROS Troponin I ES Immunodiagnostic Reagent Pack Recalled for Incorrect Wells
FDA (Devices) · 2023-08-16
- HighFDA Recalls Troponin Cardiac Test Calibrators for Potential False Negatives
FDA (Devices) · 2023-08-16
- HighOrtho Recalls VITROS Chemistry Calibrator Kit 20 for Analytical Bias
FDA (Devices) · 2023-07-19
Same hazard · measurement accuracy
See all →- HighGentuity HF-OCT imaging systems may repeat frames during catheter pullback
FDA (Devices) · 2026-04-22
- HighRadiation dosimeter may report inaccurate exposure readings
FDA (Devices) · 2023-09-20
- HighMasimo SpO2 Ear Sensor May Produce Inaccurate Oxygen Saturation Readings
FDA (Devices) · 2023-06-28
Similar recalls
- ModerateOrtho-Clinical Diagnostics Recalls VITROS Chemistry CREA Slides for Dispense Issues
FDA (Devices) · 2013-11-27
- LowOrtho-Clinical VITROS Chloride Slides Recalled for Analytical Performance Issues
FDA (Devices) · 2018-09-26
- LowOrtho-Clinical VITROS Chloride Slides Recalled for Interference Issues
FDA (Devices) · 2018-09-19
- ModerateOrtho-Clinical Diagnostics Recalls VITROS Chemistry Ca Slides Due to Dispense Issues
FDA (Devices) · 2013-11-27
- LowOrtho-Clinical Diagnostics Recalls VITROS Chemistry Products Cl- Slides
FDA (Devices) · 2018-09-19
- ModerateOrtho-Clinical VITROS UPRO Slides Recalled for Variable Urine Protein Results
FDA (Devices) · 2017-02-15