The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-3190-2017·Announced 2017-10-04

Zimmer Biomet Recalls Orthopedic Drill Bits Due to Fracture Risk

Zimmer Biomet is recalling 30 units of orthopedic drill bits because raw material anomalies may cause the instruments to fracture during use.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (instrument fracture) where injury has not yet been reported, which aligns with the 'High' severity criteria in the rubric.

Plain-English summary

Zimmer Biomet, Inc. is recalling 30 units of Drill Bit Jacobs Chuck 2.0 mm Diameter 85 mm Length (Item Number/EDI 00480508520). These nonsterile orthopedic surgical instruments were distributed nationwide in the United States and internationally to Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan, and Trinidad. The affected lot number is 63602805.

The recall was initiated because the instruments were manufactured from raw material with an anomaly. This anomaly may be on or below the surface and can increase the risk of instrument fracture. The source text does not report any specific injuries or illnesses associated with this defect.

Healthcare facilities and consumers who have these specific drill bits should stop using them immediately. Zimmer Biomet is the recalling firm responsible for managing the recall process.

The recalled product

Product
Drill Bit Jacobs Chuck 2.0 mm Diameter 85 mm Length, Item Number/EDI 00480508520, Nonsterile. orthopedic surgical instrument.
Affected units
30

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Lot Number: 63602805

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 60 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 60 products in this recall →