C.R. Bard Recalls Arctic Sun 5000 Temperature Management Systems
C.R. Bard is recalling 3,450 Arctic Sun 5000 Temperature Management Systems due to potential failure if used for more than 2,000 hours without preventive maintenance.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II and involves a risk-of-harm product where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
C.R. Bard, Inc. is recalling 3,450 units of the Arctic Sun 5000 Temperature Management System, 100-120 V North America and 100-120 V US. The recall covers Product Codes 5000-00-00 & 5000-00-00E with all serial numbers.
The recall is issued because potential failure of the system may occur when continued use exceeds 2,000 system hours without preventive maintenance. The agency has classified this as a Class II recall.
The affected units were distributed nationwide in the United States and internationally to Australia, Korea, Japan, Canada, Hong Kong, China, Singapore, and EMEA. Users should contact C.R. Bard, Inc. for instructions on how to proceed with the recall.
The recalled product
- Product
- Arctic Sun 5000 Temperature Management System, 100-120 V North America and 100-120 V US
- Manufacturer
- C.R. Bard, Inc.
- Hazard
- Affected units
- 3,450
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Product Codes 5000-00-00 & 5000-00-00E - All serial numbers
Distribution
Related recalls
Same manufacturer · C.R. Bard, Inc.
See all →- HighC.R. Bard Recalls SureStep Foley Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
- HighC.R. Bard Recalls Foley Catheter Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
- ModerateC.R. Bard Recalls Foley Catheter Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
- HighC.R. Bard Recalls SureStep Foley Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
- HighC.R. Bard Recalls SureStep Catheter Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
Same hazard · equipment failure
See all →- High
- HighHeidelberg SPECTRALIS Flex Module Recall Due to Wheel Detachment Risk
FDA (Devices) · 2026-09-30
- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22
- HighGE Optima XR200/220/240amx Radiographic System Column Safety Mechanism
FDA (Devices) · 2026-07-22
- SevereOmnipod Eros Pod Insulin Delivery Failure Risk
FDA (Devices) · 2026-07-08
Similar recalls
- HighC.R. Bard Recalls Arctic Sun Temperature Management Systems
FDA (Devices) · 2016-05-18
- HighBard Medical Recalls Arctic Sun Temperature Management Systems Due to Wiring Defect
FDA (Devices) · 2018-08-01
- SevereC.R. Bard Recalls Arctic Sun 5000 and 2000 Temperature Monitors
FDA (Devices) · 2016-02-17
- HighC.R. Bard Recalls ArcticGel Pads Due to Flow Rate Issues
FDA (Devices) · 2016-04-20
- LowC.R. Bard Recalls Arctic Sun ArcticGel Pads Due to Labeling Error
FDA (Devices) · 2017-12-13
- HighArctic Sun Temperature Management System software failure may disable critical safety alarms
FDA (Devices) · 2024-08-07