MedComp Recalls Soft-Line Duo-Flow Catheters Due to Suture Wing Cracking
Medical Components, Inc. is recalling Soft-Line Duo-Flow catheters because suture wings may crack or break, posing a risk of catheter dislocation, bleeding, or air emboli.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a defect that poses a risk of harm (bleeding, air emboli), but the source explicitly states that no serious injury or death has been reported, fitting the 'High' severity criteria for risk-of-harm products without reported injury.
Plain-English summary
Medical Components, Inc. dba MedComp is recalling Soft-Line Duo-Flow Double Lumen Catheters and Catheter Sets (models T94C, T94M, T96M, T98M). The recall is due to a defect where the suture wings have the potential to crack or break.
The company has received six complaints regarding malfunction, specifically noting 29 broken suture wings. The source text states that no serious injury or death has been reported to date. Potential patient injuries associated with this defect include catheter dislocation, bleeding, and air emboli.
The affected products were distributed nationwide and worldwide. Consumers and healthcare providers should check their inventory for the specific model numbers and lot codes listed in the official recall notice and contact the manufacturer for instructions on how to proceed.
The recalled product
- Product
- SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER/CATHETER SET; T94C T94M T96M T98M
- Manufacturer
- Medical Components, Inc dba MedComp
- Category
- Medical Device — Catheters
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 11 recalled by Medical Components, Inc dba MedComp under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 11 products in this recall →Related recalls
Same manufacturer · Medical Components, Inc dba MedComp
See all →- ModerateMedComp Recalls Power Injectable CT Port Kits with Incorrect Port Size
FDA (Devices) · 2021-03-17
- ModerateMedcomp Titan HD Catheter Kits Recalled for Incorrect Valve Size
FDA (Devices) · 2020-09-16
- SevereMedComp C3 Wave App Recalled Due to ECG Display Failure
FDA (Devices) · 2019-11-06
- ModerateMedComp Recalls Hemodialysis Catheter Kits Due to Introducer Mismatch
FDA (Devices) · 2019-08-14
- ModerateMedComp Recalls Hemodialysis Catheter Kits Due to Labeling Error
FDA (Devices) · 2019-08-14
Same hazard · catheter dislocation
See all →- HighMedComp Recalls Hemodialysis Catheter Kits Due to Suture Wing Cracking
FDA (Devices) · 2018-09-05
- HighMedComp Recalls Duo Flow Double Lumen Catheters Due to Suture Wing Cracking
FDA (Devices) · 2018-09-05
- HighMedComp Recalls Hemodialysis Catheter Kits Due to Suture Wing Cracking Risk
FDA (Devices) · 2018-09-05
- HighMedComp Duo-Flow Double Lumen Catheters Recalled for Suture Wing Cracking
FDA (Devices) · 2018-09-05
- HighMedComp Recalls Soft-Line Duo-Flow Double Lumen Catheter Trays
FDA (Devices) · 2018-09-05
Similar recalls
- SevereMedComp Recalls Catheter Repair Kits Due to Connection Failure Risk
FDA (Devices) · 2018-03-28
- HighMedComp Catheter Repair Kit Recalled for Potential Connection Failures
FDA (Devices) · 2018-03-28
- HighMedComp Recalls Tesio Catheter Sets Due to Connection Failure Risk
FDA (Devices) · 2018-03-28
- SevereMedComp Recalls 12F Tesio Catheter Sets Due to Connection Failure Risk
FDA (Devices) · 2018-03-28
- SevereMedComp Recalls 12F Modified Tesio Catheter Sets Due to Connection Failures
FDA (Devices) · 2018-03-28
- SevereMedComp Two Part Adapter Recall for Potential Connection Failures
FDA (Devices) · 2018-03-28