The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2864-2020·Announced 2020-09-02

Globus Medical Recalls Specific Lots of ALTERA Spacers Due to Material Error

Globus Medical is recalling specific lots of ALTERA Spacers because internal components may have been manufactured using stainless steel instead of the specified cobalt chromium molybdenum alloy.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II. Per the rubric, a Class II recall without explicit reports of hospitalization or injury falls under the 'High' category, representing a risk-of-harm product where injury has not yet been reported.

Plain-English summary

Globus Medical, Inc. is recalling specific lots of ALTERA Spacers. The recall involves the ALTERA Spacer, 10x31, 12-16mm, 15°.

The recall is being issued because internal components of these specific implants may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy. The affected products have been distributed worldwide.

The specific lots subject to this recall are identified by Part # 1124.1135 and the following Lot # / UDI # combinations: BAX178PE / (01)00889095069778(10)BAX178PE; BAX236NE / (01)00889095069778(10)BAX236NE; BAX398UE / (01)00889095069778(10)BAX398UE; and BAX431SE / (01)00889095069778(10)BAX431SE. Healthcare providers and patients should verify their implant lot numbers against this list and contact Globus Medical for further instructions.

The recalled product

Product
ALTERA Spacer ALTERA Spacer, 10x31, 12-16mm, 15¿

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (5)

  • Part #: 1124.1135
  • Lot # / UDI #: BAX178PE / (01)00889095069778(10)BAX178PE
  • BAX236NE / (01)00889095069778(10)BAX236NE
  • BAX398UE / (01)00889095069778(10)BAX398UE
  • BAX431SE / (01)00889095069778(10)BAX431SE

Distribution

Distribution scope not specified by the agency.

Part of a larger recall action

This product is one of 22 recalled by Globus Medical, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 22 products in this recall →

Similar recalls