Ferno iNX Wheeled Stretcher Recalled for Telescoping Load Frame Defect
Ferno-Washington Inc is recalling 72 Ferno iNX Wheeled Stretchers because hole elongation in the load frame may allow the retention pin to move.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a structural defect that could lead to patient harm, but the source text does not report any injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Ferno-Washington Inc is recalling 72 units of the Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System. The recall is due to a defect where hole elongation in the telescoping load frame channel could affect the ability of the retention pin to stay seated. This retention is necessary to preclude movement of the telescoping load frame.
The affected stretchers are used for pre-hospital patient transport. The units were distributed worldwide, including in the United States (CO, IA, IL, IN, KY, MD, MN, MO, NC, NJ, NY, OH, OK, OR, TX, VA, WA, WY) and in Canada, Australia, the United Kingdom, and New Zealand. Specific serial numbers for the affected units are listed in the official recall documentation.
Healthcare facilities and users should stop using the affected stretchers and contact Ferno-Washington Inc for instructions on how to resolve the issue. The recall is classified as Class II by the FDA.
The recalled product
- Product
- Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System Product Usage: Stretcher for pre-hospital patient transport.
- Manufacturer
- Ferno-Washington Inc
- Affected units
- 72
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial #s: 15F001703
- 15F001704
- 15F001706
- 15F001707
- 15F001708
- 15F001709
- 15F001710
- 15F001711
- 15F001712
- 15F001713
- 15F001715
- 15F001717
- 15F001719
- 15F001721
- 15F001722
- 15F001723
- 15F001724
- 15F001725
- 15F001726
- 15F001727
Distribution
Related recalls
Same manufacturer · Ferno-Washington Inc
See all →- SevereFDA Recalls Power F2 Fastening Systems Over Rail Wear Risk
FDA (Devices) · 2026-07-29
- HighPowered ambulance cots recalled because fastening post hardware could loosen
FDA (Devices) · 2025-10-22
- HighFERNO POWER X1 ambulance cot batteries pose fire and burn risk
FDA (Devices) · 2022-11-16
- ModerateFerno Chair Cot Recalled Due to Incorrect Manufacturing Nut
FDA (Devices) · 2017-04-26
- ModerateFerno-Washington Recalls Traction Splint Straps Due to Incorrect Packaging Labels
FDA (Devices) · 2016-05-11
Same hazard · structural defect
See all →- HighCanon Medical Alphenix X-ray Systems Recalled for Loose Ceiling Screws
FDA (Devices) · 2026-09-02
- HighLight Boy Recalls GBX12SK Light Towers Due to Frame Bending Risk
NHTSA · 2026-09-01
- SevereKmaier Infant Walkers Recalled for Serious Fall Hazard
CPSC · 2026-08-27
- Highrestor3d Recalls Veritas Standard Glenoid Baseplates Due to Machining Defect
FDA (Devices) · 2026-08-26
- SevereFord Recalls 2026 Maverick, F-150, and Bronco Sport for Joint Issues
NHTSA · 2026-08-25