Ferno iNX Wheeled Stretcher Recalled for Telescoping Load Frame Defect
Ferno-Washington Inc is recalling 72 Ferno iNX Wheeled Stretchers because hole elongation in the load frame may allow the retention pin to move.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a structural defect that could lead to patient harm, but the source text does not report any injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Ferno-Washington Inc is recalling 72 units of the Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System. The recall is due to a defect where hole elongation in the telescoping load frame channel could affect the ability of the retention pin to stay seated. This retention is necessary to preclude movement of the telescoping load frame.
The affected stretchers are used for pre-hospital patient transport. The units were distributed worldwide, including in the United States (CO, IA, IL, IN, KY, MD, MN, MO, NC, NJ, NY, OH, OK, OR, TX, VA, WA, WY) and in Canada, Australia, the United Kingdom, and New Zealand. Specific serial numbers for the affected units are listed in the official recall documentation.
Healthcare facilities and users should stop using the affected stretchers and contact Ferno-Washington Inc for instructions on how to resolve the issue. The recall is classified as Class II by the FDA.
The recalled product
- Product
- Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System Product Usage: Stretcher for pre-hospital patient transport.
- Manufacturer
- Ferno-Washington Inc
- Affected units
- 72
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial #s: 15F001703
- 15F001704
- 15F001706
- 15F001707
- 15F001708
- 15F001709
- 15F001710
- 15F001711
- 15F001712
- 15F001713
- 15F001715
- 15F001717
- 15F001719
- 15F001721
- 15F001722
- 15F001723
- 15F001724
- 15F001725
- 15F001726
- 15F001727
Distribution
Related recalls
Same manufacturer · Ferno-Washington Inc
See all →- SevereFDA Recalls Power F2 Fastening Systems Over Rail Wear Risk
FDA (Devices) · 2026-07-29
- HighPowered ambulance cots recalled because fastening post hardware could loosen
FDA (Devices) · 2025-10-22
- HighFERNO POWER X1 ambulance cot batteries pose fire and burn risk
FDA (Devices) · 2022-11-16
- ModerateFerno Chair Cot Recalled Due to Incorrect Manufacturing Nut
FDA (Devices) · 2017-04-26
- ModerateFerno-Washington Recalls Traction Splint Straps Due to Incorrect Packaging Labels
FDA (Devices) · 2016-05-11
Same hazard · structural defect
See all →- SevereSubaru Recalls 2026 Forester and Crosstrek for Ball Joint Defect
NHTSA · 2026-10-19
- Severe
- SevereVolvo Recalls 2025-2026 VNL Trucks for Loose Structural Bolts
NHTSA · 2026-10-15
- Severe
- Severe
Similar recalls
- HighFerno iNX Wheeled Stretcher Wheel Caster Set Screw Recall
FDA (Devices) · 2015-09-02
- HighFerno Model 35X Stretchers Recalled Due to Loose Wheel Assemblies
FDA (Devices) · 2015-01-28
- SevereFDA Recalls Power F2 Fastening Systems Over Rail Wear Risk
FDA (Devices) · 2026-07-29
- ModerateFerno-Washington Recalls Traction Splint Straps Due to Incorrect Packaging Labels
FDA (Devices) · 2016-05-11
- ModerateWinco TMM3 Video-Fluoroscopy Stretcher-Chair Recalled for Non-Conforming Component
FDA (Devices) · 2016-10-19
- HighStanding Company Recalls Superstand Wheelchairs Due to Flex Shaft Fracture Risk
FDA (Devices) · 2014-02-26