The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2801-2018·Announced 2018-08-22

ICU Medical Recalls I.V. Administration Sets Due to Leak Risk

ICU Medical is expanding a recall of I.V. Administration Sets with NanoClave manifolds due to potential leaks or component separation.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device with a structural defect (leak/separation) that poses a risk of harm, but the source text does not report any injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products without reported injury.

Plain-English summary

ICU Medical de Mexico, S.A. de C.V. is expanding the scope of a previously initiated recall to include additional I.V. Administration Sets with NanoClave manifolds. The specific product affected is the 216 cm (85") PVC/PUR Smallbore Bifuse Ext Set w/3-Port NanoClave Manifold, Check Valve, NanoClave, MicroClave Clear, Clamp, and Rotating Luer. The recall involves 50 pouches of Item Number 011-AM3010, Lot Number 3424312, with an expiration date of 03/01/2022.

The recall is being conducted due to the potential for the product to develop a leak or for separation to occur between the female luer and the NanoClave manifold. This defect could compromise the integrity of the intravenous administration system.

The affected products were distributed in the United States to Arizona, Massachusetts, North Carolina, Ohio, Texas, Virginia, and Washington. Additionally, the products were distributed worldwide to Belgium, France, and Germany. Healthcare facilities and consumers should stop using these specific lots and contact the manufacturer for further instructions.

The recalled product

Product
216 cm (85") PVC/PUR Smallbore Bifuse Ext Set w/3-Port NanoClave Manifold, Check Valve, NanoClave, MicroClave Clear, Clamp, Rotating Luer.
Affected units
50

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • Item Number 011-AM3010. Lot Number 3424312
  • UDI Number (01) 1 0840619 09487 1 (17) 220301 (30) 50 (10) 3424312
  • Expiration Date 03/01/2022

Distribution

Distributed in 7 states:

Part of a larger recall action

This product is one of 20 recalled by ICU Medical de Mexico, S.A. de C.V. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 20 products in this recall →

Same manufacturer · ICU Medical de Mexico, S.A. de C.V.

See all →

Similar recalls