Aribex Recalls NOMAD Pro2 X-Ray Systems Due to Battery Fire Risk
Aribex Inc is recalling 16,871 NOMAD Pro2 X-Ray systems because internal lithium-ion batteries may short-circuit, causing overheating, smoke, or ignition.
What this means for you
Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall involves a fire hazard (ignition) and significant injury risk (burns) from a structural defect in the battery pack, which aligns with the rubric criteria for 'Severe' (significant injury reports or structural defect).
Plain-English summary
Aribex Inc is recalling 16,871 units of the NOMAD Pro2 Packaged X-Ray System, including various part numbers and configurations such as the Vet and Black models. The recall covers units distributed across all 50 U.S. states, Puerto Rico, and numerous international locations.
The recall was initiated after the firm received customer complaints that the handsets appeared to overheat, melt, smoke, and/or ignite. An investigation determined that the internal lithium-ion batteries entered a thermal runaway condition due to a short circuit on the battery pack PCB.
Consumers should stop using the affected devices immediately and contact Aribex Inc for instructions on how to proceed with the recall.
The recalled product
- Product
- NOMAD Pro2 Packaged X-Ray System, Part Number 08500021; NOMAD Pro2 Packaged X-Ray System Black, Part Number 08500022; NOMAD Pro2 Vet Packaged X-Ray System, Part Number 08500023; NOMAD Pro2 Vet Packaged X-Ray System Black, Part Number 08500024; NOMAD Pro2 Packaged X-Ray System, GB
- Manufacturer
- Aribex Inc
- Affected units
- 16,871
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (17)
- Part Number 08500021 - Serial Number Range XD-100921-05 - XD-150109-11
- 10000 - 20811
- XD-140929-39 - XD-150123-20
- Part Number 08500022 - Serial Number Range XE-130528-03 - XE-150123-38
- 10000 - 20541
- XE-140908-32 - XE-150123-40
- Part Number 08500023 - Serial Number Range XF-121008-05 - XF-150205-03
- 10003 - 20032
- XE-150205-01
- Part Number 08500024 - Serial Number Range XG-130520-01 - XG-141210-05
- 1000 - 20019
- Part Number 08500025 - Serial Number Range XI-140303-02 - XI-150210-9
- XI-150202-04 - XI-150210-19
- Part Number 08500026 - Serial Number Range 10039 - 20041
- Part Number 08500027 - Serial Number Range 1000 - 20006
- Part Number 08500028 - Serial Number Range 10000 - 20030
- Part Number 08500029 - Serial Number Range 10000.
Distribution
Part of a larger recall action
This product is one of 2 recalled by Aribex Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Aribex Inc
See all →- SevereAribex Recalls NOMAD Pro X-Ray Systems Due to Battery Fire Risk
FDA (Devices) · 2016-09-14
Same hazard · fire
See all →- HighKibbi Recalls 2026-2027 Renegade RVs Due to Fire Risk
NHTSA · 2026-09-05
- SevereAmana Through-The-Wall Air Conditioners Recalled Due to Fire Risk
CPSC · 2026-09-03
- Severe
- High
- SevereTiffin Recalls 2026-2027 Wayfarer RVs Due to LP Tank Defects
NHTSA · 2026-08-31