The Recall Desk

FDA (Devices) recall action 74975

Aribex Inc recalled 2 products on 2016-09-14

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2016-09-14
Critical
0
Units
33,742

Why these products were recalled

Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • SevereFDA (Devices)··2016-09-14

    Aribex Recalls NOMAD Pro2 X-Ray Systems Due to Battery Fire Risk

    Aribex Inc is recalling 16,871 NOMAD Pro2 X-Ray systems because internal lithium-ion batteries may short-circuit, causing overheating, smoke, or ignition.

    Product
    NOMAD Pro2 Packaged X-Ray System, Part Number 08500021; NOMAD Pro2 Packaged X-Ray System Black, Part Number 08500022; NOMAD Pro2 Vet Packaged X-Ray System, Part Number 08500023; NOMAD Pro2 Vet Packaged X-Ray System Black, Part Number 08500024; NOMAD Pro2 Packaged X-Ray System…
    Category
    Distribution
    52 states
  • SevereFDA (Devices)··2016-09-14

    Aribex Recalls NOMAD Pro X-Ray Systems Due to Battery Fire Risk

    Aribex Inc is recalling NOMAD Pro X-Ray systems because internal lithium-ion batteries may short-circuit, causing overheating, melting, smoking, or ignition.

    Product
    NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011; NOMAD Pro Packaged X-Ray System - Europe, Part Number 08500013; NOMAD Pro X-Ray System, Remote Config., Part Number 08500017; NOMAD Pro Packaged X-Ray System JP…
    Category
    Distribution
    52 states