Mindray A7 Anesthesia Delivery System Recalled for Potential Oxygen Leak
Mindray is recalling 167 A7 Anesthesia Delivery Systems due to a potential leak on the back-up oxygen and air cylinder yokes.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a potential leak in a critical life-support device. No injuries or illnesses are reported in the source text, which aligns with the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Mindray DS USA, Inc. dba Mindray North America is recalling 167 units of the A7 Anesthesia Delivery System. This device is used to administer general inhalation anesthetics and maintain patient ventilation. The recall involves part numbers 0632F-PA0000X (US domestic) and 0632B-00014 (international).
The recall is being conducted because of a potential for a leak to occur on the back-up O2 and air e-size cylinder yokes. The affected units were distributed nationwide in the US to Alabama, Massachusetts, Maryland, Minnesota, Missouri, Nebraska, New Jersey, Oklahoma, Oregon, Pennsylvania, and Virginia, as well as to Canada.
Healthcare facilities should stop using the affected devices and contact Mindray for further instructions. The specific serial numbers for the 167 affected units are listed in the official recall documentation.
The recalled product
- Product
- A7 Anesthesia Delivery System, a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. Part numbers 0632F-PA0000X (US domestic) and 0632B-00014 (international)
- Manufacturer
- Mindray DS USA, Inc. dba Mindray North America
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- FR-5B000342
- FR-5B000343
- FR-5B000348
- FR-5B000349
- FR-51000148
- FR-51000149
- FR-51000150
- FR-51000151
- FR-51000152
- FR-51000153
- FR-51000154
- FR-51000159
- FR-51000161
- FR-51000162
- FR-51000163
- FR-51000164
- FR-51000165
- FR-51000166
- FR-51000167
- FR-51000168
Distribution
Related recalls
Same manufacturer · Mindray DS USA, Inc. dba Mindray North America
See all →- HighCentral monitoring workstations recalled over audio failure and screen freezing
FDA (Devices) · 2025-12-24
- HighAnesthesia vaporizers recalled over the potential for anesthesia leakage
FDA (Devices) · 2025-12-03
- HighIsoflurane vaporizers recalled over the potential for anesthesia leakage
FDA (Devices) · 2025-12-03
- HighPatient monitors recalled over potential activation of an abnormal alarm pause
FDA (Devices) · 2025-10-01
- HighSevoflurane Vaporizer for A9 Anesthesia System Recalled for Output Malfunction
FDA (Devices) · 2024-09-11
Same hazard · gas leak
See all →- Severe
- High
- SeverePar Nado Alto Travel Trailers Recalled for Gas Leak Risk
NHTSA · 2025-10-31
- HighHeil Recalls Side Loader Garbage Trucks Due to Fuel Leak Risk
NHTSA · 2025-10-31
- High
Similar recalls
- HighMindray A7 Anesthesia Delivery System Recalled for Software Defect
FDA (Devices) · 2018-03-14
- SevereMindray A3 Anesthesia Delivery System Recalled for Potential System Leak
FDA (Devices) · 2012-11-28
- SevereMindray A5 Anesthesia Delivery System Recalled for Potential System Leak
FDA (Devices) · 2012-11-28
- HighMindray A-Series Anesthesia Delivery System Software Recall
FDA (Devices) · 2018-03-14
- CriticalDraeger Apollo Anesthesia Machine Recalled for Potential Hypoxic Gas Delivery
FDA (Devices) · 2018-06-27
- HighIsoflurane vaporizers recalled over the potential for anesthesia leakage
FDA (Devices) · 2025-12-03