Medtronic Recalls DBS Pocket Adaptors Due to Conductor Wire Fracture Risk
Medtronic is recalling DBS Pocket Adaptors Models 64001 and 64002 due to potential conductor wire fractures. A safety notification was sent to healthcare professionals regarding device handling and integrity checks.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (conductor wire fracture in an implanted medical device) where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
Medtronic Neuromodulation is recalling the DBS Pocket Adaptor, Model 64001 and Model 64002. These devices are used as part of a neurostimulation system for deep brain stimulation and are intended to be implanted with a new replacement neurostimulator in the same pocket used for the explanted neurostimulator.
The recall is due to the potential for pocket adaptor conductor wire fractures. A safety notification was sent to healthcare professionals to reinforce device-specific labeling for the handling of pocket adaptors and system integrity checking during implant procedures.
The recall affects 7,542 units in the US and 12,999 units outside the US, with worldwide distribution. The notification is not lot specific. Healthcare professionals should refer to the safety notification for specific guidance on handling and integrity checks.
The recalled product
- Product
- DBS Pocket Adaptor, Model 64001 and Model 64002. The Medtronic Models 64001 (1x4) and 64002 (2x4) Pocket Adaptors can be used as a part of a neurostimulation system for deep brain stimulation. The pocket adaptor is intended to be implanted with the new replacement…
- Manufacturer
- Medtronic Neuromodulation
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Notification is not lot specific.
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Medtronic Neuromodulation
See all →- ModerateMedtronic 2x4 pocket adaptor kits carry incorrect use-by dates
FDA (Devices) · 2026-06-03
- ModerateMedtronic 1x4 pocket adaptor kits carry incorrect use-by dates
FDA (Devices) · 2026-06-03
- HighMedtronic SynchroMed Flex Infusion Programmer Software Application
FDA (Devices) · 2026-05-20
- ModerateNeurostimulator clinician programmer app recalled over faults during device use
FDA (Devices) · 2026-01-21
- HighFemoral nails, 360 mm left trochanteric, recalled over undersized shaft diameter
FDA (Devices) · 2025-12-31
Same hazard · wire fracture
See all →- SevereBoston Scientific Recalls RotaWire Elite Guidewires Due to Wire Fracture Complaints
FDA (Devices) · 2015-11-18
- SevereBoston Scientific Recalls RotaWire Elite Guidewires Due to Fracture Complaints
FDA (Devices) · 2015-11-18
Similar recalls
- ModerateMedtronic 1x4 pocket adaptor kits carry incorrect use-by dates
FDA (Devices) · 2026-06-03
- ModerateMedtronic 2x4 pocket adaptor kits carry incorrect use-by dates
FDA (Devices) · 2026-06-03
- HighMedtronic Recalls Deep Brain Stimulation Extension Kits Due to Impedance Reports
FDA (Devices) · 2015-05-27
- HighMedtronic Recalls 140 DBS Extension Kits with Damaged Connector Boots
FDA (Devices) · 2012-12-26
- HighMedtronic Recalls Deep Brain Stimulation Lead Kits Due to Depth Stop Defect
FDA (Devices) · 2017-08-30
- SevereMedtronic Recalls Deep Brain Stimulation Lead Kits Due to Damage Risk
FDA (Devices) · 2013-05-08