The Recall Desk
Agency classClass IOur severity:SevereFDA (Devices)·Z-2578-2020·Announced 2020-07-22

CME America Recalls CMExpress Microbore Sets Due to Delivery Inaccuracy

CME America is recalling CMExpress Microbore Sets because they may deliver fluids with significant inaccuracy, potentially causing over- or under-infusion of therapy.

The agency’s classification

FDA (Devices): Class I

Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The agency classified this as a Class I recall, and the rubric mandates a score of at least 4 for FDA Class I recalls. Additionally, the potential for significant injury due to over- or under-infusion of therapy supports a Severe rating.

Plain-English summary

CME America, LLC is recalling 113,482 units of CMExpress Microbore Sets, REF: A120-160CYFRV. The recall was initiated because the infusion administration sets may have a delivery inaccuracy exceeding plus or minus 13%, with observed ranges from minus 42.8% to plus 17.5%.

The use of these infusion pump systems could result in the over-infusion or under-infusion of medical therapy. The affected units were distributed across all 50 U.S. states, the District of Columbia, Canada, and the United Kingdom.

Healthcare facilities and consumers should stop using the affected infusion sets immediately. CME America, LLC is the recalling firm responsible for managing this safety action.

The recalled product

Product
CMExpress Microbore Sets, REF: A120-160CYFRV
Manufacturer
CME America, LLC
Affected units
113,482

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All lots within expiry

Distribution

Part of a larger recall action

This product is one of 37 recalled by CME America, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 37 products in this recall →

Same hazard · infusion inaccuracy

See all →

Similar recalls