Radiation therapy planning software can save contours to wrong image set
Elekta Monaco radiation therapy planning software has a defect allowing contour edits to be saved to the wrong image set without freezing the radiation dose for affected treatment plans.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II medical device recall affecting radiation therapy treatment planning software. Although no illnesses or injuries have been reported, the software defect presents a risk of harm by allowing treatment plan data to be saved incorrectly without proper dose-freezing controls.
Plain-English summary
Elekta Inc is recalling Elekta Monaco, a software system used to create treatment plans for patients undergoing external beam radiation therapy. The recall affects 214 units of the software worldwide, including installations in nine U.S. states (Pennsylvania, Michigan, Wisconsin, New York, Tennessee, New Jersey, Iowa, and Texas) and multiple international locations.
The software contains a defect where contour changes—outlines of anatomical structures used in treatment planning—can be inadvertently saved to an unintended image set. When this occurs, the contour edits do not cause the radiation dose to be frozen on the treatment plans associated with that image set.
Healthcare institutions using Elekta Monaco with the affected software builds (5.40.00, 5.40.01, 5.40.02, or 5.51.10) should contact Elekta Inc for corrective measures. The FDA has classified this as a Class II recall.
The recalled product
- Product
- Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions for external beam radiation therapy.
- Manufacturer
- Elekta Inc
- Affected units
- 214
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (8)
- Software Builds: 5.40.00
- 5.40.01
- 5.40.02
- 5.51.10
- UDI GTIN: (01)00858164002190(10) 5.40.00
- (01)00858164002190(10) 5.40.01
- (01)00858164002190(10) 5.40.02
- (01)00858164002268(10) 5.51.10
Distribution
Related recalls
Same manufacturer · Elekta Inc
See all →- SevereElekta Geneva Brachytherapy Applicator Intrauterine Tube Detachment Risk
FDA (Devices) · 2026-07-15
- SevereBrachytherapy Applicator Advanced Gynecological Applicator Tip Detachment
FDA (Devices) · 2026-07-15
- SevereFletcher CT/MR Shielded Applicator Set for Gynecological Brachytherapy Recalled
FDA (Devices) · 2026-07-15
- HighMOSAIQ oncology software may fail to load adapted treatment plan
FDA (Devices) · 2026-06-17
- HighLeksell Vantage Arc System locking mechanism may not secure properly
FDA (Devices) · 2026-05-06
Same hazard · software defect
See all →- HighPhilips Spectral CT Systems Recalled for Software Issues Affecting Safety
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- ModerateSteris HexaVue Focus Integration System Recalled for Patient Data Mismatch
FDA (Devices) · 2026-10-14
- ModerateUnited Imaging uMI 550 PET/CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
- ModeratePhilips SmartPath to MR 7700 Recalled for Incorrect DTI Export Data
FDA (Devices) · 2026-10-07
Similar recalls
- HighElekta Monaco RTP System Software Recall for Contour Deletion Error
FDA (Devices) · 2016-09-07
- SevereElekta Monaco Radiation Treatment Planning System Software Recall
FDA (Devices) · 2015-10-14
- SevereElekta Monaco Radiation Treatment Planning Software Recalled for Dose Errors
FDA (Devices) · 2015-11-04
- HighMonaco RTP System software may calculate radiation therapy doses inaccurately
FDA (Devices) · 2023-03-15
- HighElekta Monaco RTP System Recalled for Treatment Planning Error
FDA (Devices) · 2016-06-08
- SevereElekta MONACO RTP System Recalled for Potential Inaccurate Radiation Dose Delivery
FDA (Devices) · 2020-05-06