The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2522-2020·Announced 2020-07-15

United Imaging uCT 760 CT System Recalled for Image Artifacts

Shanghai United Imaging Healthcare is recalling five uCT 760 CT systems due to intermittent image artifacts and scanning interruptions that may require patient rescans.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (radiation exposure) where injury has not yet been reported, fitting the High severity criteria.

Plain-English summary

Shanghai United Imaging Healthcare Co., Ltd. is recalling five units of the uCT 760 System, a computed tomography x-ray system. The recall affects specific serial numbers (600229, 600230, 600232, 600271, and 600282) running specific software versions. The devices were distributed to two customers located in Houston, Texas.

The recall was initiated because two issues were identified with the system: an intermittent issue that may affect image data leading to artifacts, and intermittent scout scanning interruption caused by occasional angle signal drift. These technical failures may result in the need to rescan patients, which could expose them to an additional dose of radiation and the possible use of additional contrast media.

The source text does not specify instructions for consumers or healthcare providers on how to proceed with the recall, nor does it report any injuries or illnesses associated with the device malfunction.

The recalled product

Product
uCT 760 System, Computed Tomography X-ray System
Affected units
5

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (6)

  • Model uCT 760
  • Serial Numbers: 600229
  • 600230
  • 600232
  • 600271 with software version R001.11.0.729741-Re-1100
  • and serial number 600282 with software version R001.11.0.785930-Re-1150.

Distribution

Distributed in 1 state:

Same manufacturer · Shanghai United Imaging Healthcare Co., Ltd.

See all →

Similar recalls