The Recall Desk
HighFDA (Devices)·Z-2520-2020·Announced 2020-07-15

Kentec Medical Recalls ENFit Enteral Feeding Tubes Due to Cap Fracture Risk

Kentec Medical is recalling 170 ENFit enteral feeding tubes because the plastic tethered closure cap may fracture or break, potentially delaying patient care.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (medical device) where injury has not yet been reported, fitting the 'High' severity criteria for theoretical hazards without confirmed adverse events.

Plain-English summary

Kentec Medical, Inc. is recalling 170 units of Ameritus Medical Products ENFit Polyurethane and Silicone Feeding Tubes, catalog number ENF-Y60P-80. The specific lot number is KS1803011. These sterile, single-use tubes were distributed in California, Iowa, and Virginia.

The recall was initiated because the plastic tethered closure cap on the feeding tubes may possibly fracture or break. This defect poses two potential health hazards: first, enteral feeding and medication administration could be delayed until a replacement tube is placed; second, patients could be unnecessarily overexposed to radiation if multiple x-rays are required to confirm the placement of a new tube.

Consumers and healthcare facilities should stop using the affected feeding tubes and contact Kentec Medical for instructions on replacement or return.

The recalled product

Product
Ameritus Medical Products ENFit Polyurethane and Silicone Feeding Tubes. Sterile and single use only. Catalog number ENF-Y60P-80 (6 fr. 80 cm ENFit Enteral Feeding Tube with Y-Port).
Affected units
170

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Lot Number: KS1803011

Distribution

Distributed in 3 states:

Part of a larger recall action

This product is one of 7 recalled by Kentec Medical, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 7 products in this recall →