Inspire respiratory sensing leads labeled with Use By dates beyond shelf life
Inspire Respiratory Sensing Leads, model 4340, are being recalled because they carry an incorrect Use By Date extending beyond the validated three-year shelf life. 50 units affected.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall where an overstated expiry date could allow implantation of degraded material, with a stated risk of revision surgery and no injuries reported.
Plain-English summary
The Inspire Respiratory Sensing Lead, model/catalog number 4340, from Inspire Medical Systems, Inc., is being recalled. The action covers 50 impacted units, UDI-DI 10855728005946.
The products are labeled with an incorrect Use By Date, extending beyond the validated three-year shelf life from the date of manufacture. According to the manufacturer, material degradation of a product past its Use By Date may result in unplanned or premature revision surgery. The FDA classified the action as Class II.
Distribution was worldwide, including nationwide in the United States and to Hong Kong. The leads are supplied to healthcare facilities rather than sold to consumers.
Facilities should identify affected stock using the model number, UDI-DI and part numbers in the recall notice, quarantine it, and follow the instructions issued by Inspire Medical Systems.
The recalled product
- Product
- Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a…
- Manufacturer
- Inspire Medical Systems Inc.
- Affected units
- 50
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Lot Code: Model No 4340 (USA only lead)
- UDI-DI [4340] 10855728005946
- Inspire part number [4340] 900-014-003
- serial numbers T18806
- T45012
- T45017
- T49002
- T67098
- T67152
- T69541
- T69542
- T69553
- T69558
- T69562 Model No 4340 (Global lead)
- UDI-DI [4340] 10810098650301
- Inspire part number [4340] 900-014-005
- serial numbers T91485
- T91486
- T91491
- T91495
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 2 recalled by Inspire Medical Systems Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Inspire Medical Systems Inc.
See all →- HighInspire stimulation leads labeled with Use By dates beyond shelf life
FDA (Devices) · 2026-06-24
- ModerateInspire Model 4063 Stimulation Lead Labeling Error — Serial Number Mismatch
FDA (Devices) · 2024-10-16
- ModerateInspire Respiratory Sensing Lead Model 4340 Serial Number Label Mismatch
FDA (Devices) · 2024-10-16
- SevereInspire Model 3028 Implantable Pulse Generator Manufacturing Defect
FDA (Devices) · 2024-07-03
- ModerateInspire II Implantable Pulse Generator Recalled for Incorrect Use-By Date
FDA (Devices) · 2018-09-12
Same hazard · labeling error
See all →- ModerateLiquid Hope Organic Meal Replacement Formula Subpotent Vitamin A Recall
FDA (Food) · 2026-07-22
- ModerateOrthofix MINIRAIL external fixation devices recalled for incorrect model labeling
FDA (Devices) · 2026-07-22
- Moderate
- HighPerampanel Prescription Tablets Recalled for Labeling Mix-up
FDA (Drugs) · 2026-07-01
- HighAdndale magnesium glycinate gummies contain undeclared melatonin in three lots
FDA (Food) · 2026-06-24
Similar recalls
- ModerateInspire Respiratory Sensing Lead Model 4340 Serial Number Label Mismatch
FDA (Devices) · 2024-10-16
- ModerateInspire Model 4063 Stimulation Lead Labeling Error — Serial Number Mismatch
FDA (Devices) · 2024-10-16
- ModerateInspire II Implantable Pulse Generator Recalled for Incorrect Use-By Date
FDA (Devices) · 2018-09-12
- HighAbbott Infinity 5 Pulse Generator May Have Inaccurate Battery-Life Labels
FDA (Devices) · 2024-07-31
- ModerateBEAR implants recalled over an expiration date beyond approved shelf life
FDA (Devices) · 2026-02-04
- HighAbbott Proclaim 5 Elite implantable pulse generator service life may be shorter than labeled
FDA (Devices) · 2024-07-24