The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2476-2026·Announced 2026-06-24

Giraffe OmniBed Carestation canopy fasteners can fall into patient compartment

Giraffe OmniBed Carestations are being recalled because fasteners securing the canopy soffit can disengage and fall into the patient compartment. 23,430 units affected.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: FDA Class II recall with a stated risk of patient injury from parts falling into an infant's compartment, a risk of harm with no injuries reported.

Plain-English summary

Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination infant incubator and infant warmer, model numbers 2082844-001-XXX, UDI-DI 010084068211686221, are being recalled. The action covers 23,430 units, identified by the serial numbers in the recall notice.

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example impact during transport, one or more of the fasteners that secure the soffit can disengage, and a fastener could fall into the patient compartment. According to the manufacturer, in rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach, and these conditions could potentially result in patient injury. The FDA classified the action as Class II.

Distribution was worldwide, including nationwide in the United States. The devices are installed in healthcare facilities rather than sold to consumers.

Facilities should identify affected units using the model number, UDI-DI and serial numbers in the recall notice and follow the instructions issued by GE HealthCare.

The recalled product

Product
Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination of an infant incubator and an infant warmer, Model Numbers 2082844-001-XXX
Affected units
23,430

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • UDI-DI: 010084068211686221

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 3 recalled by GE Medical Systems, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 3 products in this recall →