The Recall Desk
HighFDA (Devices)·Z-2476-2026·Announced 2026-06-24

[pending] Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination of an infant incubat

Pending LLM rewrite. Source: FDA_DEVICE Z-2476-2026.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Pending LLM scoring against the rubric.

Plain-English summary

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

The recalled product

Product
Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination of an infant incubator and an infant warmer, Model Numbers 2082844-001-XXX
Manufacturer
GE Medical Systems, LLC

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • UDI-DI: 010084068211686221

Distribution

Distributed nationwide across the United States.