The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2455-2015·Announced 2015-09-02

Omnilife Recalls Offset Femoral Alignment Guides Due to Incorrect Markings

Omnilife Science Inc. is recalling 15 Offset Femoral Alignment Guides because the markings are reversed, which may prevent proper alignment during surgery.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device with a defect that could lead to improper surgical alignment, representing a risk of harm where injury has not yet been reported, consistent with the 'High' severity criteria.

Plain-English summary

Omnilife Science Inc. is recalling 15 units of the Offset Femoral Alignment Guide 4mm x 6 (Product Code: KS 67024). The recall affects specific lot numbers: MM0314R, VA0713, and Z1111. These instruments are designed to facilitate bone preparation and implantation for the Apex Knee or Apex Hip Systems.

The reason for the recall is that the markings on both versions of the alignment guides are incorrect. Specifically, the markings are reversed from the intended position, which should provide proper alignment for offset revision stems when used with the femur.

The affected products were distributed worldwide, including in the United States (specifically in Oklahoma, Colorado, Ohio, and Massachusetts) and in Spain and Australia. Healthcare providers and facilities that have received these products should stop using them and contact the recalling firm for instructions on return or replacement.

The recalled product

Product
Offset Femoral Alignment Guide 4mm x 6¿; Product Code: KS 67024 Product Usage: The instrumentation has been designed to facilitate bone preparation and implantation of the Apex Knee or Apex Hip Systems. Product Usage: The instrumentation has been designed to facilitate…
Affected units
15

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • Lot Numbers: MM0314R
  • VA0713
  • Z1111

Distribution

Distributed nationwide across the United States.

The notice also names 4 states specifically:

Part of a larger recall action

This product is one of 2 recalled by Omnilife Science Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 2 products in this recall →