Biomerieux VITEK 2 Gram Negative Susceptibility Card Container Integrity Recall
Biomerieux Inc is recalling VITEK 2 Gram Negative Susceptibility cards because compromised container integrity may expose the product to moisture, potentially causing antibiotic degradation and loss of potency.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves potential loss of potency in a medical diagnostic device, which poses a risk of harm through inaccurate results, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
Biomerieux Inc is recalling 2,359 cartons of VITEK 2 Gram Negative Susceptibility cards (AST-N270, REF 414162). The recall was initiated because the integrity of the product container may be compromised, which could expose the product to moisture. This exposure may lead to antibiotic degradation and a loss of potency.
The affected product was distributed nationwide in the United States, including to military and government sites. The product was also distributed to numerous international locations, including Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroun, Canada, Chad, Chile, China, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Gambia, Georgia, Germany, Greece, Grenada, Guatemala, Guinea, Haiti, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Uganda, Kuwait, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Malta, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Zimbabwe.
Healthcare facilities and distributors should stop using the affected product and contact Biomerieux Inc for further instructions. The recall is classified as Class II by the FDA.
The recalled product
- Product
- VITEK¿ 2 Gram Negative Susceptibility card (AST-N270), REF 414162, 20 cards per carton.
- Manufacturer
- Biomerieux Inc
- Affected units
- 2,359
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- 6900190203 04/Jun/18
Distribution
Part of a larger recall action
This product is one of 187 recalled by Biomerieux Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 187 products in this recall →Related recalls
Same manufacturer · Biomerieux Inc
See all →- HighVITEK 2 AST-N447 Test Kit Potentially Returns False Resistant Results
FDA (Devices) · 2026-07-15
- HighSusceptibility test cards recalled over false resistant antibiotic results
FDA (Devices) · 2026-02-11
- HighAntimicrobial susceptibility test cards recalled over false colistin resistance results
FDA (Devices) · 2025-10-08
- SevereVITEK 2 AST Test Kits: Ceftriaxone Concentration Error Risk
FDA (Devices) · 2024-05-01
- HighMedical Device Middleware: MYLA Software Fails to Apply AST Filter Rules
FDA (Devices) · 2023-03-08
Same hazard · loss of potency
See all →- ModerateInfuvite Adult Multiple Vitamins Injection Recalled Due to Cold Chain Failure
FDA (Drugs) · 2025-05-07
- ModerateCardinal Health Recalls Cyanocobalamin Injection Due to Temperature Excursions
FDA (Drugs) · 2021-06-02
- ModerateSmiths Medical Recalls Arterial Blood Sampling Kits with Povidone Iodine Pads
FDA (Devices) · 2019-02-27
- ModerateBiomerieux Recalls VITEK 2 Gram Negative Susceptibility Cards Due to Moisture Risk
FDA (Devices) · 2017-06-21
- ModerateBiomerieux Recalls VITEK 2 Gram Negative Susceptibility Cards
FDA (Devices) · 2017-06-21