COPAN FLOQSwabs Recalled Over Sterilization Integrity Failure
Copan Italia is recalling 3,029,400 COPAN FLOQSwabs distributed across the US due to sterilization integrity failures. The clinical sampling swabs cannot be guaranteed sterile due to data integrity issues at the contract sterilizer.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall involving sterile medical devices where sterility cannot be guaranteed due to documented integrity issues. While no injuries or illnesses have been reported, the affected devices present a risk of harm as medical devices intended for clinical sampling.
Plain-English summary
COPAN FLOQSwabs are sterile medical devices used for clinical sample collection. Copan Italia is recalling 3,029,400 units that were distributed nationwide in the US and Puerto Rico.
The recall was initiated because a sterility assurance level of 10-6 cannot be guaranteed. This is due to intentional fraud and data integrity issues at Steril Milano, the contract sterilizer. These issues mean the manufacturer cannot verify that the affected devices meet the required sterility standards.
The recalled product
- Product
- COPAN FLOQSwabs, STERILE.EO, for drawing clinical samples. CE0123, for the following reference numbers REF:501CS01, Minitip, sterile Single Wrapped, Molded bp 80mm; REF: 20226, P/N 501CS01Q100, Minitip, Sterile, Single Wrapped, bp 60MM, BUNDLES OF 20 pcs; REF…
- Manufacturer
- Copan Italia
- Affected units
- 3,029,400
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Mode Number: Lot number: 501CS01: 1816581 and 2027890
- 501CS01.Q100: PJ10453
- PJ10636 and PJ10667
- 501CS01.Q20: J10407
- 502CS01: 1816964
- 2027239
- 2027877
- 503CS01: 1807589
- 1809095
- 1814761
- 1814762
- 1814763
- 1816968
- 1821362
- 2025887
- 503CS01.BD: 2031801
- 503CS01.BX: 50U004DS and 1818685
- 50U009DS02: 1825274
- 518CS01: 2008058 and 2027915
- 518CS01.BD: 2008058
Distribution
Related recalls
Same manufacturer · Copan Italia
See all →- ModerateSputum transfer reagent sets recalled because liquification cannot be guaranteed
FDA (Devices) · 2025-12-24
- ModerateRespiratory specimen reagent recalled because fluidification cannot be guaranteed
FDA (Devices) · 2025-12-24
- ModerateBD BBL CultureSwab EZ Collection and Transport System Swabs Breakage
FDA (Devices) · 2025-06-18
- LowCopan WASP Instrument Recalled for Software Upgrade to Prevent Table Rotation
FDA (Devices) · 2013-09-11
Same hazard · sterilization failure
See all →- SevereSteris VERIFY Biological Indicator Recall Due to Spore Population Concerns
FDA (Devices) · 2026-10-14
- HighSteris VERIFY Dual Species Biological Indicator Recall for Spore Population Issues
FDA (Devices) · 2026-10-14
- SevereSteris VERIFY Biological Steam Test Pack Recalled for Spore Population Defect
FDA (Devices) · 2026-10-14
- SevereSteris VERIFY Biological/Integrator Challenge Pack Recall for Spore Population Defect
FDA (Devices) · 2026-10-14
- SevereSteris Recalls VERIFY Biological Steam Test Packs Due to Spore Population Concerns
FDA (Devices) · 2026-10-14
Similar recalls
- HighCodan Swan-Lock Needlefree Connectors Recalled for Potential Sterility Compromise
FDA (Devices) · 2020-07-08
- HighCopan Transfer Pipette Recalled Due to Plastic Defects
FDA (Devices) · 2023-09-27
- SevereCustomed Recalls Surgical Packs Due to Potential Packaging Integrity Issues
FDA (Devices) · 2014-09-10
- SevereCustomed Recalls Cesarean Packs Due to Potential Packaging Integrity Issues
FDA (Devices) · 2014-09-10
- SevereCustomed Recalls Laparotomy Packs Due to Potential Packaging Integrity Issues
FDA (Devices) · 2014-09-10
- SevereCustomed Recalls EO Sterilized Surgical Packs Due to Packaging Integrity Concerns
FDA (Devices) · 2014-09-10