Suntech Oscar 2 Blood Pressure Monitor Recall for Display Error
Suntech Medical is recalling Oscar 2 blood pressure monitors because the display may show an incorrect decimal point for three-digit readings in mmHg.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the device is a medical instrument where incorrect readings pose a risk of harm, although no specific injuries or illnesses have been reported in the source text.
Plain-English summary
Suntech Medical, Inc. is recalling 1,037 units of the Oscar 2, Model 250 System blood pressure monitor. The recall was initiated because the device display can show an incorrect decimal point when the blood pressure reading is a three-digit number and the unit of measure is set to mmHg.
The affected units were distributed nationwide to California, Connecticut, Florida, Illinois, Louisiana, Maryland, North Carolina, New York, Ohio, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, and Utah. They were also distributed worldwide to Australia, Austria, Belgium, Bermuda, Brazil, Canada, Croatia, Czech Republic, Dominican Republic, Estonia, France, India, Ireland, Latvia, Lebanon, Morocco, Pakistan, Peru, Poland, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, United Kingdom, and Vietnam.
Consumers should check their device serial number against the list provided in the official recall notice to determine if their unit is affected. If the serial number matches, users should stop using the device and contact Suntech Medical for instructions on repair or replacement.
The recalled product
- Product
- Oscar 2, Model 250 System
- Manufacturer
- Suntech Medical, Inc.
- Affected units
- 1,037
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Catalog Numbers (Serial Numbers): 99-0133-00 (SN:00101086
- SN:00101159
- SN:00101248
- SN:00102173
- SN:00107798
- SN:00112277
- SN:00113151
- SN:00113262
- SN:00113268
- SN:00113279
- SN:00113717
- SN:00113838
- SN:00114084
- SN:00114513
- SN:00115988
- SN:00116238
- SN:00116258
- SN:00116331
- SN:00116380
- SN:00116384
Distribution
Related recalls
Same manufacturer · Suntech Medical, Inc.
See all →- HighSuntech Medical Recalls Model 250D Ambulatory Blood Pressure Monitors for Battery Safety
FDA (Devices) · 2020-01-22
- ModerateSunTech Medical Recalls Mislabeled Neonate Blood Pressure Cuffs
FDA (Devices) · 2019-08-21
Same hazard · display error
See all →- ModerateKico Knee Innovation Recalls ARVIS Surgical Navigation System Due to Sensor Issue
FDA (Devices) · 2026-08-26
- ModerateSpectral Instruments Recalls 35 Ami HTX X-Ray Units
FDA (Devices) · 2026-08-26
- ModerateSpectral Instruments Lago X Medical Device Recalled for Display Glitch
FDA (Devices) · 2026-08-26
- ModerateLinkBio CORE Workstation may display an incorrect planning date
FDA (Devices) · 2026-04-22
- HighHexagon Purus Recalls Tern RC8 Vehicles Due to Brake Display Error
NHTSA · 2026-04-20