Sedecal RadPro 40kW Digital X-Ray Panel Recalled for Uncontrolled Movement Risk
Sedecal S.A. is recalling RadPro 40kW Digital X-Ray systems because electrostatic energy may cause uncontrolled movements that could bump into users or patients.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (uncontrolled movement causing physical impact) where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
Sedecal S.A. is recalling the RadPro 40kW Digital X-Ray Panel and Mobile/Portable Diagnostic X-Ray systems. The recall involves 361 systems in the United States and 5 systems in foreign markets, distributed nationwide in the US and in Canada. The devices are intended for use by qualified doctors or technicians for diagnostic radiographic exposures.
The recall is due to a potential hazard where electrostatic energy may be stored in the systems. This stored energy can cause a short-circuit, which may generate an electromagnetic pulse. Such a pulse could disturb the electronics of the Digital Motion Control Board, producing uncontrolled movements. These uncontrolled movements could cause the system to bump into the user or patient.
Affected units are identified by specific serial numbers listed in the recall documentation. Users should contact Sedecal S.A. for further instructions on how to address this issue.
The recalled product
- Product
- RadPro 40kW Digital X-Ray Panel and Mobile/Portable Diagnostic X-Ray. Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and…
- Manufacturer
- Sedecal S.A.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial numbers: G29393
- G29395
- G29404
- G29416
- G29426
- G29864
- G29873
- G29878
- G30025
- G30155
- G30156
- G30160
- G30165
- G30173
- G30236
- G30240
- G30250
- G30254
- G30258
- G30567
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Sedecal S.A.
See all →- SevereSedecal NOVA FA DR System Recalled for Potential Cable Failure
FDA (Devices) · 2018-11-28
- HighSedecal Mobile X-ray Systems Recalled for Uncommanded Movements
FDA (Devices) · 2016-03-16
- HighSedecal MobileDiagnost wDR X-ray systems recalled for unexpected movement
FDA (Devices) · 2015-02-11
Same hazard · uncontrolled movement
See all →- HighRebstock Holding Rod for Fixation May Move Uncontrollably
FDA (Devices) · 2024-03-06
- HighSkytron GS70H Surgical Table Recall: Control Failure Causes Continued Movement
FDA (Devices) · 2023-09-06
- HighMedical Device Recall: X-Ray Detector Drive Shaft May Break
FDA (Devices) · 2023-06-07
- HighX-Ray Machine Detector May Move Unexpectedly Due to Drive Shaft Failure
FDA (Devices) · 2023-06-07
- Highda Vinci Surgical System Patient Cart Ballscrew Failure Risk
FDA (Devices) · 2023-03-22
Similar recalls
- HighSedecal Mobile X-ray Systems Recalled for Uncommanded Movements
FDA (Devices) · 2016-03-16
- HighSedecal MobileDiagnost wDR X-ray systems recalled for unexpected movement
FDA (Devices) · 2015-02-11
- HighVirtual Imaging RadPRO Mobile X-Ray System Recalled for Mount Separation Risk
FDA (Devices) · 2018-11-21
- ModerateFour 40KW 50C X-ray systems recalled with water resistance reminder
FDA (Devices) · 2025-10-29
- HighSedecal Mobile X-ray System Recalled for Software Defect
FDA (Devices) · 2016-06-15
- ModerateForty kilowatt 55G X-ray systems recalled with water resistance reminder
FDA (Devices) · 2025-10-29