Integra Recalls 2.7mm Drilling Guides Due to Breakage During Use
Integra Life Sciences is recalling 2,389 units of 2.7mm Drilling Guides because they may break during surgical use, potentially increasing procedure time.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The source explicitly states that no patient injuries resulted from the breakage, and the identified risk is limited to increased surgical time, fitting the criteria for a moderate severity recall involving low-risk issues without reported harm.
Plain-English summary
Integra Life Sciences is recalling 2,389 units of 2.7mm Drilling Guides, Catalogue No. 219635ND. These reusable instruments are used in conjunction with a drill to prepare oriented holes in bone for the insertion of locking screws in various compression plate systems, including Tiblaxis, Uni-CP, Advansys MLP, and Advansys DLP.
The recall was initiated after Integra's quality system identified an adverse trend in complaints regarding the breakage of the drilling guides during use. While none of the reported complaints resulted in patient injuries, the breakage in some cases led to an increase in the time required to complete the surgical procedure. A physician consulted to assess patient risk concluded that any potential risk would be transient, self-limiting, or minor.
The affected units were distributed worldwide, including nationwide in the United States and in several countries in Europe, the Middle East, and Africa. Specific lot numbers for the affected guides are listed in the official recall documentation. Users should stop using the affected devices and contact the manufacturer for instructions on replacement or return.
The recalled product
- Product
- The 2.7mm Drilling Guide is a reusable instrument used in conjunction with a drill to prepare oriented holes in bone for insertion of a Standard Surfix¿ Locking Screw with plates for the Tiblaxis, Uni-CP, Advansys MLP and Advansys DLP compression plate systems. The Drilling…
- Manufacturer
- Integra Life Sci.
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- The Lot No's for the affected 2.7mm Drilling Guides are: E1DT
- E4EX
- E57T
- E5YA
- E64W
- E6GN
- E6RB
- E6WP
- E7UL
- E86B
- E8UC
- E9PX
- E9Q5
- E9QD
- EBNY
- EEYJ
- EF11
- EFCM
- EGT6
- EH6X
Distribution
Related recalls
Same hazard · breakage
See all →- ModeratePapablic Replacement Brush Heads Recalled for Potential Breakage
FDA (Devices) · 2026-07-22
- ModeratePapablic Replacement Brush Heads Recalled Due to Breakage Risk
FDA (Devices) · 2026-07-22
- SevereMedline Convenience Kits with Recalled Swan-Ganz Catheters
FDA (Devices) · 2026-07-08
- HighMedline Convenience Kits with Recalled Swan-Ganz Catheters
FDA (Devices) · 2026-07-08
- HighMedline Convenience Kits Containing Recalled Swan-Ganz Catheters
FDA (Devices) · 2026-07-08
Similar recalls
- HighMedtronic Prestige LP Cervical Disc Drill Guides Recalled for Breakage
FDA (Devices) · 2018-05-16
- HighSynthes Pull Reduction Device Recalled for Potential Intraoperative Breakage
FDA (Devices) · 2015-06-24
- HighSurgical drill guide may bend and produce metal shavings during use
FDA (Devices) · 2023-07-12
- HighIntegra Advansys Medial Lisfranc Plate Recalled for Surgical Technique Risk
FDA (Devices) · 2012-08-08
- HighSurgical drill guides may bend and produce metal shavings during use
FDA (Devices) · 2023-07-12
- LowIntegra NeuraGen Nerve Guide Recalled for Quality Control Error
FDA (Devices) · 2020-03-18