The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2363-2019·Announced 2019-09-04

Abbott Recalls Alinity ci Series System Control Modules Due to Software Error

Abbott is recalling Alinity ci series System Control Modules because a software issue may cause incorrect test results after a system stop.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (medical device) where incorrect results could impact patient care, but no injuries or illnesses have been reported in the source text.

Plain-English summary

Abbott GmbH & Co. KG is recalling 1,763 units of the Alinity ci series System Control Module, Part Number 03R70 01. The recall covers all serial numbers and includes all on-market software versions v2.6.0 and v2.6.1. These modules are used in clinical chemistry and immunoassay analysis systems.

The recall was initiated because a software issue may cause incorrect results to occur after a system Stop. This specific issue only happens if the system is transitioned from Processing to Stopped to Idle, which leads to the Alinity i reusing reaction vessels. The product was distributed worldwide and nationwide in the US.

Healthcare facilities and users of the Alinity ci series should contact Abbott for instructions on how to address the software issue. No specific consumer action is required for the general public, as this is a medical device used in professional settings.

The recalled product

Product
Alinity ci -series System Control¿Module, Part Number 03R70 01 Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a…
Affected units
1,763

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All serial numbers

Distribution

Distributed nationwide across the United States.

Similar recalls