LINK MP Monoblock Hip Stem trial instruments recalled for increased removal difficulty
Waldemar Link GmbH recalled LINK MP Monoblock hip stem trial instruments (Size 23) nationwide due to increased force required for removal post-impaction, potentially prolonging surgery time.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II hip implant trial instrument recall involving a risk-of-harm product where increased removal difficulty may extend surgical duration. This documented issue during surgical procedures meets the criteria for High severity classification.
Plain-English summary
Waldemar Link GmbH & Co. KG is recalling LINK MP Monoblock Hip Stem trial instruments (Size 23) distributed nationwide in the United States. The issue affects all sizes of this trial stem product (14mm to 25mm).
The trial instruments may require increased force to remove after impaction during surgery. This can result in prolonged surgical time or require modification of the planned surgical procedure.
These instruments are intended for hip joint procedures. The increased removal force needed may extend the duration of surgery and complicate the surgical procedure.
Healthcare providers using these instruments should be aware of the potential for increased removal force during procedures.
The recalled product
- Product
- LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 23 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-23/00
- Manufacturer
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Category
- Medical Device — Hip Implant
- Affected units
- 24
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI-DI: 04026575182176 Lot Numbers: B922118 B933169 C005023 C211017 C240048
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 12 recalled by Waldemar Link GmbH & Co. KG (Mfg Site) under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 12 products in this recall →Related recalls
Same manufacturer · Waldemar Link GmbH & Co. KG (Mfg Site)
See all →- HighWaldemar Link Recalls Endo-Model Tibial Plateau Replacement Sets
FDA (Devices) · 2026-08-19
- ModerateEmbrace standard drill tower surgical technique lists mixed-up article numbers
FDA (Devices) · 2026-04-01
- ModerateEmbrace shoulder drill tower surgical technique lists mixed-up article numbers
FDA (Devices) · 2026-04-01
- HighLink Endo-Model plateau 15-0027/15 recalled over screw bushing fracture
FDA (Devices) · 2026-03-18
- HighLink Endo-Model SL component 16-2840/05 recalled over bushing fracture
FDA (Devices) · 2026-03-18
Same hazard · removal difficulty
See all →- HighMedtronic Percutaneous Surgical Pins Recalled Due to Potential Fitting Defects
FDA (Devices) · 2024-01-17
- HighMedtronic Percutaneous Pins Recalled for Potential Fitting and Removal Issues
FDA (Devices) · 2024-01-17
- HighMedical Device Recall: LINK MP Hip Stem Trial Instruments Removal Difficulty Risk
FDA (Devices) · 2023-08-16
- HighLINK MP Hip Stem trial instruments may require increased removal force during surgery
FDA (Devices) · 2023-08-16
- ModerateHip Stem Trial Instrument Difficult to Remove After Impaction
FDA (Devices) · 2023-08-16
Similar recalls
- HighWaldemar Link Recalls Trial Stems for Hip Reconstruction Prosthesis
FDA (Devices) · 2013-09-04
- ModerateHip Reconstruction System Component Mislabeled for Size
FDA (Devices) · 2025-01-01
- ModerateWaldemar Link Recalls Trial Stems for Reconstruction Prosthesis
FDA (Devices) · 2013-09-04
- HighDePuy SIGMA HP MBT Non-Keel Punch Knee Instrument Recalled
FDA (Devices) · 2016-02-17
- HighZimmer M/L Taper Hip Stem Recall Due to Black Residue
FDA (Devices) · 2014-07-16
- ModerateHip Prosthetic Stem Recalled for Incorrect Product Code Label
FDA (Devices) · 2025-03-26