RayStation 9B Software Recalled for Dose Calculation Error
RaySearch Laboratories is recalling RayStation 9B software due to a dose calculation error that could underestimate radiation therapy doses.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (radiation therapy software) where a dose underestimation could lead to treatment failure, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
RaySearch Laboratories AB is recalling RayStation 9B software, identified by UDI # 07350002010266 and build number 9.1.0.933. This software is used for radiation therapy and medical oncology to administer treatments. The recall is due to a software error identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms.
The error could result in a local underestimation of the expected dose during treatment. The affected software was distributed nationwide in the United States, including in Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.
Healthcare providers using this software should contact RaySearch Laboratories for further instructions on how to address the dose calculation error.
The recalled product
- Product
- RayStation 9B, UDI # 07350002010266 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI # 07350002010266 Build number: 9.1.0.933
Distribution
Part of a larger recall action
This product is one of 15 recalled by RAYSEARCH LABORATORIES AB under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 15 products in this recall →Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · dose calculation error
See all →- HighRadiation Therapy Planning System Density Calculation Inconsistency in RayStation
FDA (Devices) · 2025-02-26
- HighRayStation radiation therapy planning software density uncertainty calculation inconsistency
FDA (Devices) · 2025-02-26
- HighMedical device recall: RayStation radiation therapy software dose calculation inconsistency
FDA (Devices) · 2025-02-26
- HighRayStation Radiation Therapy Planning Software Recall for Density Calculation Inconsistency
FDA (Devices) · 2025-02-26
- HighRadiation Therapy Planning Software Correction for Density Uncertainty in Dose Calculations
FDA (Devices) · 2025-02-26
Similar recalls
- HighRayStation Radiation Therapy Software Recalled for Incorrect Dose Calculation
FDA (Devices) · 2023-07-05
- HighRaySearch Recalls Radiation Therapy Planning Software Due to Dose Calculation Errors
FDA (Devices) · 2019-07-10
- SevereRaySearch Ray Station 4.0 software recalled for dose calculation errors
FDA (Devices) · 2014-04-09
- HighRayStation Radiation Therapy Software Recalled for Dose Calculation Accuracy Issues
FDA (Devices) · 2018-07-04
- HighRayStation Software Recall for Radiation Therapy Dose Calculation Errors
FDA (Devices) · 2014-06-25
- HighRadiation therapy planning system dose calculation error
FDA (Devices) · 2024-05-08