RayStation 6 Radiation Therapy Software Recalled for Dose Calculation Error

RaySearch Laboratories is recalling RayStation 6 software due to a calculation error that may underestimate radiation dose.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (radiation therapy software) where a dose underestimation could lead to treatment failure or injury, but no specific injuries or illnesses are reported in the source text.

Plain-English summary

RaySearch Laboratories AB is recalling RayStation 6 software, identified by UDI # 07350002010013 and Build number 6.0.0.24. This software is used for radiation therapy and medical oncology to administer treatments.

The recall was initiated because a software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms. This error could result in a local underestimation of the expected dose. The product was distributed nationwide in the United States, including in Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.

Healthcare providers and facilities using this software should contact RaySearch Laboratories for further instructions on how to address the calculation error.

The recalled product

Product
RayStation 6, UDI # 07350002010013 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • UDI # 07350002010013 Build number: 6.0.0.24

Distribution

Distributed nationwide across the United States.