Agency classClass IIOur severity:HighFDA (Devices)·Z-2310-2019·Announced 2019-08-28

BD Vacutainer Lithium Heparin Tubes Recalled for False COHb Results

Becton Dickinson is recalling BD Vacutainer LH blood collection tubes because they may cause false elevation of carboxyhemoglobin results when used with the IL GEM 4000 instrument.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II. The hazard involves diagnostic inaccuracy (false elevation of COHb) which poses a risk of harm through potential misdiagnosis or inappropriate treatment, but no specific injuries or illnesses are reported in the source text.

Plain-English summary

Becton Dickinson & Company is recalling BD Vacutainer LH (Lithium Heparin) 102 I.U. Plus Blood Collection Tubes, Catalog Number 368886. The recall involves 1,436,228,004 units that were distributed worldwide, including nationwide in the United States.

The recall was initiated because false elevation of carboxyhemoglobin (COHb) results have been reported when venous blood gas samples collected with these tubes are analyzed using the IL GEM 4000 instrument. This issue may affect healthcare facilities and laboratories using the specific combination of these collection tubes and the IL GEM 4000 analyzer.

Healthcare providers and laboratories should stop using the affected tubes for venous blood gas samples intended for analysis on the IL GEM 4000 instrument. Users should contact Becton Dickinson & Company for further instructions regarding the return or replacement of the product.

The recalled product

Product
BD Vacutainer LH (Lithium Heparin) 102 I.U. Plus Blood Collection Tubes, Catalog Number 368886
Affected units
1,436,228,004

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • UDI: (01)30382903666646

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 15 recalled by Becton Dickinson & Company under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 15 products in this recall →