Elekta Unity Software Recall for Motion Monitoring Data Mismatch
Elekta Limited is recalling specific Unity software versions due to a potential mismatch between contour data and motion monitoring images.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II and involves a risk-of-harm product (medical device software) where the hazard is a data mismatch, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
Elekta Limited is recalling specific versions of Elekta Unity and Philips Marlin software (R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, and R6.1.571). The recall involves 26 units distributed in the United States to Iowa, New York, Pennsylvania, Texas, and Wisconsin, as well as internationally to Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, the Netherlands, Sweden, Switzerland, Turkey, and the United Kingdom.
The recall was initiated because, in Treatment Session Manager (TSM) Motion Monitoring workflows, there can be a mismatch between the contour data overlay and the motion monitoring images of the monitored structure under certain conditions. This issue affects specific anatomy protocols including Abdomen, HeadNeck L, HeadNeck M, Pelvis L, Pelvis M, and Thorax.
The FDA has classified this as a Class II recall. Users of the affected software should contact Elekta Limited for further instructions on how to resolve the issue.
The recalled product
- Product
- Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571
- Manufacturer
- Elekta Limited
- Affected units
- 26
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (19)
- Software versions: R5.3.30
- R5.3.31
- R6.0.5331
- R5.7.0
- R5.7.1
- btFFE Cor Sag RealTime
- btFFE Cor Sag Tra RealTime
- btFFE Fast Sag Cor RealTime
- btFFE Sag RealTime)
- HeadNeck L (bFFE Cor RealTime
- bFFE Cor Sag RealTime
- bFFE Cor Sag Tra RealTime
- bFFE Sag RealTime
- bFFE Sag Tra RealTime)
- HeadNeck M (bFFE Cor RealTime
- Pelvis L (bFFE Cor RealTime
- Pelvis M (bFFE Cor RealTime
- Thorax (btFFE Cor RealTime
- btFFE Sag RealTime
Distribution
Related recalls
Same manufacturer · Elekta Limited
See all →- ModerateElekta Unity Systems Recalled for Missing Contrast Agent Warnings
FDA (Devices) · 2019-05-15
- HighElekta Unity MRI Radiation Therapy System Recalled for Image Protocol Issues
FDA (Devices) · 2019-04-17
- HighElekta Unity MRI Speakers May Become Dislodged and Enter Bore
FDA (Devices) · 2019-01-16
- HighElekta Integrity R1.2 Medical Device Recalled for Positional Error Risk
FDA (Devices) · 2018-03-07
- HighElekta Agility with Integrity R3.2 Software Recall for Positional Errors
FDA (Devices) · 2018-03-07
Same hazard · software defect
See all →- HighPhilips Spectral CT Systems Recalled for Software Issues Affecting Safety
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- ModerateSteris HexaVue Focus Integration System Recalled for Patient Data Mismatch
FDA (Devices) · 2026-10-14
- ModerateUnited Imaging uMI 550 PET/CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
- ModeratePhilips SmartPath to MR 7700 Recalled for Incorrect DTI Export Data
FDA (Devices) · 2026-10-07
Similar recalls
- ModerateElekta Unity Radiation Therapy System Software Recall
FDA (Devices) · 2019-07-24
- HighElekta Unity MRI Radiation Therapy System Recalled for Image Protocol Issues
FDA (Devices) · 2019-04-17
- HighElekta Focal SIM Planning Software Recalled for Incorrect Patient Shift Directions
FDA (Devices) · 2012-07-18
- ModerateElekta Unity Systems Recalled for Missing Contrast Agent Warnings
FDA (Devices) · 2019-05-15
- HighElekta Unity Radiation Therapy System Recalled for Heating Issues
FDA (Devices) · 2020-07-08
- HighElekta Agility software recall for potential calibration input errors
FDA (Devices) · 2014-03-12