Maquet Recalls TEGRIS System Due to Missing Emergency Stop Button
Maquet is recalling 21 TEGRIS systems because a secondary touch screen display lacks an emergency stop button.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a missing safety feature (emergency stop button) where no injuries or illnesses have been reported, fitting the rubric for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Maquet Cardiovascular US Sales, LLC is recalling 21 units of the TEGRIS system. The recall is due to a potential issue where the secondary touch screen display lacks an emergency stop button.
The TEGRIS system is a central operating system for operating rooms that records and distributes images and videos and interacts with medical and non-medical devices. It is operated using a touchscreen and can control devices such as operating room tables and lights. The affected units are the Tegris Base System (1703.01X0) and OR Tablet Module (1703.71A0) used in combination with the Secondary 19" touchscreen (part number 1707.05A0).
The affected units were distributed worldwide, including in the United States (New York and Pennsylvania) and internationally to Australia, Brazil, Canada, Germany, Hong Kong, Japan, and Sweden. The source text does not specify instructions for consumers or healthcare providers on how to proceed with the recall.
The recalled product
- Product
- TEGRIS system Four labels of the Tegris System include: Emergency button (1703.72) MAQUET GmbH & Co. KG Kehler-Strabe 31 Baden-Wuerttemberg 76437 Rastatt, Germany Touch screen (elo) Elo TouchSystems 301 Constitution Drive Menlo Park, CA 94025…
- Manufacturer
- Maquet Cardiovascular Us Sales, Llc
- Hazard
- Affected units
- 21
Distribution
Related recalls
Same manufacturer · Maquet Cardiovascular Us Sales, Llc
See all →- HighMaquet Cardiovascular Recalls HLS Set Advanced Due to Sterility Breaches
FDA (Devices) · 2020-12-23
- HighMaquet Cardiovascular Recalls HLS Set Advanced Due to Sterility Breach
FDA (Devices) · 2020-12-23
- HighMaquet BMU 40 Blood Monitoring Units Recalled for Electrical Safety Non-Compliance
FDA (Devices) · 2020-11-18
- HighMaquet Yuno II Mobile Operating Table Recalled for Electric Shock Risk
FDA (Devices) · 2020-10-28
- HighMaquet Yuno Mobile Operating Table Recalled for Electric Shock Risk
FDA (Devices) · 2020-10-28
Same hazard · missing safety feature
See all →- High
- HighMaterial processing lasers recalled for lacking a required remote interlock connector
FDA (Devices) · 2026-01-07
- High
- HighJaguar Recalls 2022 F-Pace Vehicles for Missing Audio Safety Features
NHTSA · 2022-08-04
- ModerateZiehm Vision C-Arm X-Ray Systems Recalled for Emergency Switch Issue
FDA (Devices) · 2014-01-01
Similar recalls
- HighMaquet Tegris System Recalled Due to Software Error in Operating Table Commands
FDA (Devices) · 2015-01-28
- ModerateMaquet CARDIOHELP Base Unit Recalled for Display and Alarm Defects
FDA (Devices) · 2013-11-06
- HighMaquet Extension Device Recalled for Potential Detachment from Operating Table
FDA (Devices) · 2016-03-09
- HighMaquet Cardiovascular Recalls Custom Tubing Sets for Extracorporeal Circulation
FDA (Devices) · 2019-06-05
- HighMaquet CARDIOHELP System extracorporeal support device components recalled for sterility
FDA (Devices) · 2023-03-08
- HighMaquet Getinge Custom Tubing Sets Recalled for Sterility Assurance Concerns
FDA (Devices) · 2019-06-05