The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2236-2021·Announced 2021-08-18

Orthopedic drill bit recalled for corrosion particle contamination risk

NEOSTEO cannulated drill bits are being recalled due to corrosion traces on the laser marking that could release particles into the surgical field. The recall affects Lot 243-20 distributed in New Jersey.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: FDA Class II medical device recall. The source does not report any illnesses or injuries to date. However, the detected corrosion defect presents a risk-of-harm in a surgical context (potential for inflammation, allergic reaction, and adverse tissue response), which elevates this to High severity despite the absence of reported incidents.

Plain-English summary

NEOSTEO is recalling Cannulated Graduated Drill Bits (1.9 mm diameter, REF: INS-114) due to traces of corrosion detected on the laser marking. The affected lot is 243-20, distributed in New Jersey. This surgical instrument is used in orthopedic procedures to prepare bone for screw placement.

The corrosion on the laser marking could cause wear and corrosion particles to be released during the drilling procedure. If these particles contact bone and adjacent tissue during surgery, they could potentially cause inflammation, pain, allergic reaction, or adverse tissue response.

Healthcare providers and surgical facilities should immediately remove drill bits from Lot 243-20 from service. Contact NEOSTEO regarding proper return procedures.

The recalled product

Product
Cannulated graduated drill bit 1.9 mm diameter, with AO QC system, REF: INS-114 for use in orthopedic surgery.
Manufacturer
NEOSTEO
Affected units
25

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Lot: 243-20

Distribution

Distributed nationwide across the United States.

The notice also names 1 state specifically:

Similar recalls