The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2215-2024·Announced 2024-07-03

CentriMag Circulatory Support Console May Shut Down During Power Surge

CentriMag Primary Console units may completely shut down without alarm during power surges above 1.8kV, requiring emergency switch to backup circulatory support.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: This is a Class II FDA recall of a critical medical device component. The failure mode—complete shutdown without alarm during power surge conditions—creates serious risk to patients dependent on circulatory support. The hazard is conditional, triggered only by specific power surge levels (1.8–2.0kV). Per FDA Class II classification and the rubric criteria for risk-of-harm medical devices, this scores as High severity.

Plain-English summary

Thoratec Switzerland GMBH is recalling the 2nd Generation CentriMag Primary Console, part of the CentriMag Circulatory Support System. Approximately 3,425 consoles have been distributed worldwide, including throughout the United States and to numerous countries internationally.

A component change in the console may cause it to fail to conform to the IEC 61000-4-5 electrical surge standard. When exposed to a power surge above 1.8 kilovolts and up to 2.0 kilovolts, the affected console may shut down completely without triggering an alarm. When this occurs, the pump ceases functioning and the patient must be switched to a backup circulatory support system.

The recalled product

Product
2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411, L201-90421, part of the CentriMag Circulatory Support System
Affected units
3,425

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (8)

  • 07640135140726
  • 07640135140702
  • 07640135140719
  • 07640135140856
  • 05415067037282
  • 07640135140894
  • 07640135149019
  • 07640135149002

Distribution

Distributed nationwide across the United States.

The notice also names 37 states specifically:

Same manufacturer · Thoratec Switzerland GMBH

See all →