Hycor Biomedical Recalls HYTEC 288 System and RF Assay Kits
Hycor Biomedical is recalling HYTEC 288 System and RF assay kits because high IgG levels may cause false positive results, leading to incorrect rheumatoid diagnoses.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (medical device) where the hazard is an incorrect diagnosis, but no specific injuries or illnesses have been reported in the source text, fitting the 'High' severity criteria for risk-of-harm products without reported injury.
Plain-English summary
Hycor Biomedical Inc is recalling the HYTEC 288 System, Model 74007P, and associated HYTEC AUTOSTAT II RF assay kits (IgM, IgA, IgG, and Total Rheumatoid Factor). The recall covers all lots of the RF assays. The products were distributed worldwide to the United States (Texas), Netherlands, Belgium, Austria, Italy, and India.
The recall is due to reports that control and negative samples may show elevated results following a sample containing high levels of IgG. This issue can cause negative controls to report out of range or negative samples to be recovered as low positives, which could contribute to an incorrect or premature diagnosis of rheumatoid arthritis.
Healthcare facilities and laboratories using these instruments and kits should stop using the affected products and contact Hycor Biomedical Inc for instructions on returning or replacing the equipment and kits.
The recalled product
- Product
- HYTEC 288 System, Model 74007P, automated immunoanalyzer (EIA) instrument. For in vitro diagnostic use only. HYTEC AUTOSTAT II RF assays: HYTEC AUTOSTAT II RF IgM Kit: Part Number FGA 05, HYTEC AUTOSTAT II RF IgA Kit: Part Number FGA 06, HYTEC AUTOSTAT II RF IgG Kit: Part Nu
- Manufacturer
- Hycor Biomedical Inc
- Affected units
- 288
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All lots of RF assay
Distribution
Distributed in 1 state:
Related recalls
Same manufacturer · Hycor Biomedical Inc
See all →- ModerateHycor Biomedical Recalls AUTOSTAT II IgM Rheumatoid Factor Test Kits
FDA (Devices) · 2019-08-07
Same hazard · false positive
See all →- HighHealthmark ProChek-II Medical Tests Recalled for Expired Ingredient
FDA (Devices) · 2026-09-02
- HighAmyloid 1-42-N plasma calibrators recalled over falsely elevated results
FDA (Devices) · 2026-02-11
- HighpTau 217 plasma cartridges recalled over falsely elevated ratio results
FDA (Devices) · 2026-02-11
- HighpTau 217 plasma calibrators recalled over falsely elevated ratio results
FDA (Devices) · 2026-02-11
- HighAmyloid 1-42-N plasma cartridges recalled over falsely elevated results
FDA (Devices) · 2026-02-11