The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2196-2012·Announced 2012-08-22

Edwards Lifesciences Recalls Swan-Ganz Catheters Due to Guidewire Passage Issues

Edwards Lifesciences is recalling Swan-Ganz Thermodilution VIP Catheters because a J-tip guidewire may pass through the distal lumen hub.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II, and the source text describes a risk-of-harm product defect (guidewire passage) without explicitly reporting injuries or illnesses, which aligns with the 'High' severity criteria.

Plain-English summary

Edwards Lifesciences, LLC is recalling its Swan-Ganz Thermodilution VIP Catheters, Model 151F7. These devices are used for hemodynamic monitoring in patients. The recall involves 32,145 units distributed worldwide, including the United States, Canada, Japan, and various regions in Europe, Asia, and Latin America.

The recall was initiated after the firm received multiple complaints regarding the passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter. This specific mechanical issue is the reason for the corrective action.

Healthcare facilities and distributors should stop using the affected catheters and contact Edwards Lifesciences for further instructions. The specific product codes associated with this recall are 59174304, 59174305, 59174306, 59196997, 59211405, 59211406, 59211407, 59211408, 59216088, 59233788, 59233789, 59253934, 59252521, 59252522, 59252523, and 59252524.

The recalled product

Product
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 151F7. For use in patients who require hemodynamic monitoring.
Affected units
32,145

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (16)

  • 59174304
  • 59174305
  • 59174306
  • 59196997
  • 59211405
  • 59211406
  • 59211407
  • 59211408
  • 59216088
  • 59233788
  • 59233789
  • 59253934
  • 59252521
  • 59252522
  • 59252523
  • 59252524

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 11 recalled by Edwards Lifesciences, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 11 products in this recall →