Stryker Target Detachable Coils Recalled for Manufacturing Defect
Stryker Neurovascular is recalling Target Detachable Coils because manufacturing damage to internal sutures may reduce stretch resistance, potentially leading to increased coil stretching.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a structural defect in a medical device that could lead to device malfunction (coil stretching) without reported injuries, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Stryker Neurovascular is recalling specific units of Target Detachable Coils, including the Target Helical Nano 2mm x 4cm model. The recall affects 45,302 units worldwide, with 8,188 units distributed in the United States. The product is intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
The recall was initiated because some units do not have the intended stretch resistance. Stryker Neurovascular identified that the sutures inside the coil, which provide this resistance, may have been damaged during the manufacturing process. This defect can lead to an increased occurrence of coil stretching.
The affected products were distributed worldwide, including in the United States, Australia, Canada, and numerous other countries. Specific lot numbers and expiration dates are listed in the agency's recall record. Healthcare providers and patients should contact Stryker Neurovascular for further instructions on how to proceed with the recall.
The recalled product
- Product
- Target Detachable Coils: TARGET HELICAL NANO 2MM X 4CM MODEL Number:M0035452040 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
- Manufacturer
- Stryker Neurovascular
- Category
- Medical Device — Neurovascular
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- MODEL Number:M0035452040
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 28 recalled by Stryker Neurovascular under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 28 products in this recall →Related recalls
Same manufacturer · Stryker Neurovascular
See all →- HighStryker Neurovascular INZONE Detachment System battery drain
FDA (Devices) · 2026-07-01
- HighTrevo Trak 21 Microcatheter Distributed with Incorrect Instructions and Labels
FDA (Devices) · 2024-10-16
- HighStryker Synchro Neuro Guidewires Recalled for PTFE Coating Damage Risk
FDA (Devices) · 2024-06-12
- HighSynchro SELECT Guidewires may have PTFE coating damage
FDA (Devices) · 2024-06-12
- HighSynchro2 Guidewires: PTFE coating damage from improper backloading technique
FDA (Devices) · 2024-06-12
Same hazard · device malfunction
See all →- SevereMedtronic MiniMed Flex Insulin Pumps Recalled for Failed Verification Tests
FDA (Devices) · 2026-10-14
- ModerateWilson-Cook Recalls Hemospray Endoscopic Hemostat Due to Adherence Issues
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Et Control Software Recalled for Anesthetic Delivery Interruption
FDA (Devices) · 2026-09-23
Similar recalls
- ModerateStryker Neurovascular Recalls Expired Target Detachable Coils
FDA (Devices) · 2014-12-10
- ModerateStryker GDC 360 Coils Recalled for Expiration Date Violations
FDA (Devices) · 2014-12-10
- ModerateStryker GDC 360 Detachable Coils Recalled for Incorrect Device Formulation
FDA (Devices) · 2016-04-27
- ModerateStryker Neurovascular Microcatheters Recalled for Incorrect Use-By Dates
FDA (Devices) · 2013-12-25
- ModerateStryker Neurovascular Microcatheters Recalled for Shelf-Labeling Errors
FDA (Devices) · 2013-12-25
- ModerateStryker Neurovascular Recalls Expired Gateway PTA Balloon Catheters
FDA (Devices) · 2014-12-10