RingLoc Hip Acetabular Cup Recalled for Assembly Defect
Biomet recalled RingLoc hip acetabular cup components that may have been incorrectly assembled with a mismatched locking ring, potentially causing joint instability. Affected patients may require surgical intervention.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II device with a structural assembly defect (mismatched locking ring) that poses risk of joint instability requiring surgical intervention. No injuries or hospitalizations have been reported; this is a preemptive recall for a potential hazard.
Plain-English summary
Biomet, Inc. is recalling RingLoc Bi-Polar Hip System Acetabular Cup components (41 mm outer diameter, item number 11-165206) used in hip hemi-arthroplasty. The recall affects 12 units with distribution in the United States (Arizona, California, New Jersey, Pennsylvania, Texas, and Virginia) and Chile.
The recalled units may have been incorrectly assembled with a 42 mm locking ring instead of the correct size. This mismatch between the 41 mm acetabular cup and the 42 mm locking ring could cause joint instability in the implanted hip.
Joint instability may require surgical intervention to address. Patients who have received hip implants using affected components should contact their surgeon or physician immediately. The recalled lot numbers are: 66540443, 66601057, 66623486, and 66635486.
The recalled product
- Product
- RingLoc Bi-Polar Hip System Acetabular Cup, 41 mm Outer Diameter - Used in hip hemi-arthroplasty Item Number: 11-165206
- Manufacturer
- Biomet, Inc.
- Affected units
- 12
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- Lot Number / UDI Number (1) 66540443 (01)00880304001923(17)290408(10)66540443
- (2) 66601057 (01)00880304001923(17)290405(10)66601057
- (3) 66623486 (01)00880304001923(17)290327(10)66623486
- (4) 66635486 (01)00880304001923(17)290327(10)66635486
Distribution
Related recalls
Same manufacturer · Biomet, Inc.
See all →- ModerateIntramedullary Bone Saw Blade Assembly Recalled Due to Potential Packaging Damage
FDA (Devices) · 2026-08-12
- HighIntramedullary Bone Saw Blade Assemblies Recalled Due to Potential Packaging Damage
FDA (Devices) · 2026-08-12
- ModerateIntramedullary Bone Saw Blade Assembly Recalled Due to Potential Sterile Barrier Damage
FDA (Devices) · 2026-08-12
- HighIntramedullary Bone Saw Blade Assembly Recalled Due to Potential Sterile Barrier Damage
FDA (Devices) · 2026-08-12
- HighIntramedullary Bone Saw Blade Assembly Recall for Packaging Damage
FDA (Devices) · 2026-08-12
Same hazard · assembly defect
See all →- ModerateOctopus Evolution Tissue Stabilizer Model TS2000 tubing assembly error
FDA (Devices) · 2026-07-01
- ModerateOctopus 4 Tissue Stabilizer incorrectly assembled tubing positions
FDA (Devices) · 2026-07-01
- ModerateOctopus Evolution AS Tissue Stabilizer Model TS2500 Tubing Assembly Issue Recall
FDA (Devices) · 2026-07-01
- ModerateCardiosave battery charging station left bay screw prevents full battery insertion
FDA (Devices) · 2026-03-25
- HighMedtronic tissue stabilizer recalled after tubing was assembled in reverse
FDA (Devices) · 2026-03-04
Similar recalls
- HighBiomet Recalls RingLoc Hip System Components Due to Endotoxin Levels
FDA (Devices) · 2020-06-03
- HighBiomet Recalls RingLoc Hip Cups Due to Double Sterilization
FDA (Devices) · 2020-10-21
- ModerateBiomet Recalls Acetabular Shell Components Due to Packaging Mix-Up
FDA (Devices) · 2015-02-25
- ModerateBiomet Recalls Specific Lot of Mallory Head Acetabular Shell Components
FDA (Devices) · 2015-02-25
- HighBiomet Hip Prosthesis Recall for Potential Sterile Seal Breach
FDA (Devices) · 2020-12-23
- ModerateBiomet Recalls Active Articulation E1 Hip Bearing Implants Due to Labeling Errors
FDA (Devices) · 2013-06-12