Siemens SHBG Immulite 2000 Reagent Recall Due to Stability Issues
Siemens is recalling certain lots of SHBG Immulite 2000 reagents because they may not meet stability claims, potentially causing test bias.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves potential test bias without reported injuries or illnesses, fitting the criteria for a moderate severity voluntary precautionary recall.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling 9,100 kits of Sex Hormone Binding Globulin (SHBG) Immulite 2000/IMMULITE 2000 XPi reagents. The affected lots are 357, 358, 360, 361, and 363, with expiration dates ranging from January 30, 2017, to June 30, 2017. These products were distributed nationwide.
The recall was initiated because certain lots of the SHBG reagents do not meet the 30-day open vial adjustor stability claim when reconstituted and stored at 2-8°C, as stated in the Instructions for Use. If the adjustors are stored at 2-8°C for more than 14 days, they may show a bias in patient samples and quality control results may fall outside of published ranges.
The FDA has classified this as a Class II recall. The source text states that a lookback is not recommended. Users should discontinue use of the affected lots and contact the manufacturer for further instructions.
The recalled product
- Product
- SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Affected units
- 9,100
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (7)
- Lot Numbers: 1) 357
- 2) 358
- 3) 360
- 4) 361 & 5) 363 Expiration Dates: 1) 2017-01-30
- 2) 2017-03-31
- 3) 2017-04-30
- 4) 2017-05-31 & 5) 2017-06-30
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · stability failure
See all →- ModerateSafecor Health Recalls Fluphenazine HCl Elixir Due to Stability Failures
FDA (Drugs) · 2026-09-23
- ModerateSafecor Health Recalls Fluphenazine HCl Elixir Due to Stability Failures
FDA (Drugs) · 2026-09-23
- ModerateSafecor Health Recalls Fluphenazine HCl Elixir Due to Stability Failures
FDA (Drugs) · 2026-09-23
- ModerateSafecor Health Recalls Fluphenazine HCl Elixir Due to Stability Failures
FDA (Drugs) · 2026-09-23
- ModerateSafecor Health Recalls Fluphenazine HCl Elixir Due to Stability Failures
FDA (Drugs) · 2026-09-23
Similar recalls
- HighSiemens Recalls IMMULITE Assays Due to Biotin Interference Risk
FDA (Devices) · 2018-05-16
- HighSiemens IMMULITE Thyroid Test Kit Recalled for Falsely Low Results
FDA (Devices) · 2022-10-05
- ModerateSiemens Recalls IMMULITE 2000 Total IgE Assays Due to Positive Bias
FDA (Devices) · 2014-05-07
- ModerateSiemens Recalls Intact PTH Assay Kits Due to Negative Bias
FDA (Devices) · 2017-01-18
- ModerateSiemens IMMULITE Total IgE Assay Recalled for Positive Bias
FDA (Devices) · 2014-05-07
- HighSiemens IMMULITE TBG Test Kits Recalled for Inaccurate Results
FDA (Devices) · 2015-05-27